D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 9 Modifier Addition<br />

Figure 9-2 Two pump modifier system<br />

Note<br />

Make sure you purchase a “dip-tube” CO 2 tank.<br />

To fill pump A with<br />

supercritical fluid (helium<br />

head-space tank)<br />

1. Place the pumps in independent control - constant pressure<br />

mode, as detailed above.<br />

2. Be sure the pumps are stopped.<br />

3. Close the pump A outlet valve before opening the tank supply<br />

valve and pump A inlet valve.<br />

4. Press the REFILL key on the front panel keypad. Press<br />

softkey A to designate that pump A is the active pump (the<br />

pump being refilled). Pump A will fill with pressurized<br />

fluid.<br />

9-6