Section 9 Modifier Addition<br />

9.3 Initial Preparation for<br />

Modifier Addition<br />

When first setting up the D <strong>Series</strong> pump system under modifier<br />

mode, the pumps should be zeroed and filled. To accomplish this,<br />

the pumps should first be placed under independent control -<br />

constant pressure mode.<br />

Note<br />

Once the modifier system is operating, the system does not<br />

need to be zeroed before refilling.<br />

To designate independent<br />

control (constant pressure<br />

mode)<br />

To designate constant<br />

pressure<br />

To zero the pumps<br />

To fill pump B with modifier<br />

1. If the pumps are running, press the STOP key once.<br />

2. Press the MENU key.<br />

3. Press softkey A - MORE.<br />

4. Press 4 - MULTI-PUMP, to display the multi-pump menu.<br />

5. Press 4 - INDEPENDENT to designate that the pumps be<br />

operated under independent control. The number 4 will<br />

blink, indicating that independent has been selected.<br />

6. Press softkey D - PREVIOUS, twice to return to the main<br />

menu.<br />

1. Press the blue CONST PRESS key on the front panel keypad.<br />

1. Place the pumps in independent control - constant pressure<br />

mode, as detailed above.<br />

2. Be sure the pumps are stopped.<br />

3. Disconnect the inlet tubing from the inlet valve of pump A.<br />

4. Select 2 pump independent control/constant pressure<br />

mode.<br />

5. Open the inlet valves of each pump.<br />

6. Run both pumps to their empty positions.<br />

7. Zero the pressure on each pump.<br />

8. Reconnect the pump A inlet tubing to the inlet valve.<br />

1. Place the pumps in independent control - constant pressure<br />

mode, as detailed above.<br />

2. Be sure the pumps are stopped.<br />

3. Close the pump B outlet valve and open the pump B inlet<br />

valve.<br />

4. Press the REFILL key on the front panel keypad. Press<br />

softkey B to designate that pump B is the active pump (the<br />

pump being refilled). Pump B will fill with modifier.<br />

5. After pump B has filled, press run and deliver 5-10 ml of<br />

modifier back through the inlet valve, to ensure all air is<br />

purged from the pump, before pressing stop.<br />

6. Close the inlet valve.<br />

9-5