07.03.2014 Views

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 9 Modifier Addition<br />

9.1 Introduction The Teledyne <strong>Isco</strong> modifier addition systems utilize two D <strong>Series</strong><br />

syringe pumps regulated by one controller to deliver a mixture of<br />

supercritical fluid and modifier to one or more Model SFX 2-10,<br />

SFX 220, or SFX 3560 extractors.<br />

Note<br />

The configurations described in this section apply to 65DM,<br />

100DM, 100DX, 260D, 500D, and 1000D pumps that use Valco<br />

or SSI fittings.<br />

The 65D pump can be configured and operated in a similar<br />

manner. However, due to the higher operating pressure and<br />

the AE F250C fittings, the 65D requires hardware that is not<br />

listed in this section. Call the factory for details to configure a<br />

65D pump for this type of operation.<br />

Pump A delivers supercritical fluid while pump B meters modifier.<br />

Both fluids pass through one-way check valves before they<br />

meet and are blended together in a mixing tee. Modifier concentrations<br />

of up to 100 percent (v/v) are programmable via the controller.<br />

If higher concentrations are needed, put the modifier in<br />

pump A to be delivered as the primary fluid.<br />

Both pumps operate in constant pressure mode and will flow up<br />

to their maximum rate to maintain the set pressure. When modifier<br />

addition mode is selected and the RUN key is pressed, pump<br />

A pressure is ramped at its maximum flow rate to the set<br />

pressure while pump B waits at a lower pressure (3.5 bar).<br />

The user can increase the pump B minimum pressure if the modifier<br />

boils at room temperature and 3.5 bar. See Section 9.4, Modifier<br />

Addition.<br />

Once pump A has reached the set pressure, the controller will<br />

enter a hold mode and wait for a pressure drop, with opening the<br />

extractor inlet to signal the start of “EQUILIBRATION,” or the<br />

user can press the “RUN” key to start equilibration.<br />

Note<br />

Flow must then be started through the system by opening all<br />

valves! The modifier pump will not equilibrate properly without<br />

flow.<br />

Then pump B pressure is slowly ramped upward until its check<br />

valve is detected to open. At this point, the controller exits<br />

“EQUILIBRATION” mode and displays the message<br />

“RUNNING” on the front panel display.<br />

9-1

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!