D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 8 Continuous Flow Introduction, Installation, & Operation<br />

8.6.2 Continuous Flow<br />

Features<br />

Volume Totalizer<br />

To reset the volume totalizer<br />

to zero<br />

Valve Specification<br />

The following features are used to define your continuous flow<br />

operation.<br />

The total volume delivered (in liters) is given on the display at<br />

the right top corner.<br />

1. Press the orange MENU key on the controller front panel.<br />

2. Press softkey A, MORE, to access the second page of the<br />

main menu.<br />

3. Press 5 on the numeric keypad to select 5. TOTAL VOL-<br />

UME RESET.<br />

4. Press softkey D, PREVIOUS, twice to return to the run<br />

screen. The displayed volume will be zero.<br />

It is important to specify in the menu the type of valve package<br />

being used for continuous flow. This will assure minimum<br />

pressure fluctuation at switchover.<br />

To specify valves installed<br />

8.7 Operating Tips and<br />

Guidelines<br />

Operational Overview –<br />

When using liquids<br />

1. Press the orange MENU key on the controller front panel.<br />

2. Press softkey A, MORE, to access the second page of the<br />

main menu.<br />

3. Press 6 on the numeric keypad to select 6. VALVE.<br />

4. Press 1 or 2 on the numeric keypad to specify Active (air<br />

actuator) or Passive (check valves) installed.<br />

5. Press softkey D, PREVIOUS, three times to return to the<br />

run screen.<br />

The following tips and guidelines are provided by our research<br />

laboratory. Please familiarize yourself with them before operating<br />

your continuous flow system.<br />

• Temperature changes can cause pressure fluctuations,<br />

especially if a restrictor is being used for backpressure.<br />

• Pressure limits for continuous constant flow mode are<br />

set by the limits of pump A.<br />

• For correct overpressure response, overpressure<br />

shutdown must be set to shutdown “ON” under pump<br />

limit options.<br />

• Always check the flow rate before pressing the RUN key,<br />

especially after using the pump in a mode other than<br />

continuous flow. For example, a flow rate can be carried<br />

over from independent mode control of pump A.<br />

• Refill rates: As a rule of thumb, the refill rate should be<br />

at least twice the flow rate setpoint (in continuous<br />

constant flow) to allow time for refill and repressurization<br />

before the next switchover. The same refill rate<br />

should be entered separately for pumps A and B.<br />

1. Degas liquids if appropriate.<br />

2. Fill both pumps completely.<br />

8-18