Section 8 Continuous Flow Introduction, Installation, & Operation<br />
8.6.2 Continuous Flow<br />
Features<br />
Volume Totalizer<br />
To reset the volume totalizer<br />
to zero<br />
Valve Specification<br />
The following features are used to define your continuous flow<br />
operation.<br />
The total volume delivered (in liters) is given on the display at<br />
the right top corner.<br />
1. Press the orange MENU key on the controller front panel.<br />
2. Press softkey A, MORE, to access the second page of the<br />
main menu.<br />
3. Press 5 on the numeric keypad to select 5. TOTAL VOL-<br />
UME RESET.<br />
4. Press softkey D, PREVIOUS, twice to return to the run<br />
screen. The displayed volume will be zero.<br />
It is important to specify in the menu the type of valve package<br />
being used for continuous flow. This will assure minimum<br />
pressure fluctuation at switchover.<br />
To specify valves installed<br />
8.7 Operating Tips and<br />
Guidelines<br />
Operational Overview –<br />
When using liquids<br />
1. Press the orange MENU key on the controller front panel.<br />
2. Press softkey A, MORE, to access the second page of the<br />
main menu.<br />
3. Press 6 on the numeric keypad to select 6. VALVE.<br />
4. Press 1 or 2 on the numeric keypad to specify Active (air<br />
actuator) or Passive (check valves) installed.<br />
5. Press softkey D, PREVIOUS, three times to return to the<br />
run screen.<br />
The following tips and guidelines are provided by our research<br />
laboratory. Please familiarize yourself with them before operating<br />
your continuous flow system.<br />
• Temperature changes can cause pressure fluctuations,<br />
especially if a restrictor is being used for backpressure.<br />
• Pressure limits for continuous constant flow mode are<br />
set by the limits of pump A.<br />
• For correct overpressure response, overpressure<br />
shutdown must be set to shutdown “ON” under pump<br />
limit options.<br />
• Always check the flow rate before pressing the RUN key,<br />
especially after using the pump in a mode other than<br />
continuous flow. For example, a flow rate can be carried<br />
over from independent mode control of pump A.<br />
• Refill rates: As a rule of thumb, the refill rate should be<br />
at least twice the flow rate setpoint (in continuous<br />
constant flow) to allow time for refill and repressurization<br />
before the next switchover. The same refill rate<br />
should be entered separately for pumps A and B.<br />
1. Degas liquids if appropriate.<br />
2. Fill both pumps completely.<br />
8-18