D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 8 Continuous Flow Introduction, Installation, & Operation<br />

8.4 Continuous Flow<br />

Electric Valves<br />

Electric valves, which provide the best positive valve closure, are<br />

most commonly used in liquefied gas applications. Electric valves<br />

require the pump controller to be equipped with the valve driver<br />

board.<br />

Note<br />

The max flow rate for the E1000 dual system will tend to have<br />

a lower percentage flow rate due to port size restrictions, with<br />

a maximum flow rate equal to that of the E500 system<br />

(see Table 8-5).<br />

Electric Valve Package<br />

Numbers<br />

E1000 68-1247-109<br />

E500 68-1247-091<br />

E260, E100M, E100X 68-1247-090<br />

Note<br />

Electric valves are not available for the 65D pump. Please consult<br />

the factory for options.<br />

8.4.1 Technical<br />

Specifications<br />

Dual system maximum flow rates under optimal conditions are<br />

approximately 65% of the max flow rate for each pump model.<br />

Table 8-5 Continuous Flow Technical Specifications-Electric Valves:<br />

Pressure fluctuation at switchover<br />

Minimum<br />

Maximum system backpressure<br />

Maximum flow rate (ml/min)<br />

Liquids: 65% of the single pump<br />

maximum rate<br />

(excluding E1000 systems).<br />

Liquefied gases<br />

0.35 bar, at system backpressures from 6.9 bar to the single<br />

pump maximum. Higher fluctuation occurs at pressures below<br />

6.9 bar.<br />

3.5 bar<br />

Temperature range 0 to 40°C<br />

Wetted materials in valve packages:<br />

Check valves<br />

Tubing and fittings<br />

The single pump maximum. Valves rated to 689.5 bar<br />

E1000 132.6<br />

E500 132.6<br />

E260 69.55<br />

E100X 32.50<br />

E100M 16.25<br />

E65DM 19.50<br />

45% of the single pump maximum rate. Cylinder cooling jackets<br />

should be used to obtain this rate.<br />

SS316, sapphire, ruby, PEEK, PTFE<br />

SS316, gold<br />

8-13