07.03.2014 Views

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 8 Continuous Flow Introduction, Installation, & Operation<br />

8.3 Continuous Flow Air<br />

Valves<br />

Air Valve Package<br />

Numbers<br />

Air valves generally open and close faster and are less prone to<br />

error and component wear, making them suitable for industrial<br />

use and other applications where the system will be constantly<br />

running; however, they require a user-supplied pressurized air<br />

source of 80 to 115 psi (5.5 to 7.9 bar).<br />

A1000 68-1247-104<br />

A500 68-1247-061<br />

A65DM, A260, A100M, A100X 68-1247-058<br />

A65 68-1247-129<br />

Table 8-4 Continuous Flow Technical Specifications-Air Valves:<br />

Pressure fluctuation at switchover<br />

Minimum<br />

Maximum system backpressure<br />

Air supply source pressure<br />

Maximum flow rate (ml/min)<br />

Liquids: 65% of the single pump<br />

maximum rate.<br />

Liquefied gases<br />

5 psi (0.35 bar), at system backpressures from 100 psi<br />

(6.9 bar) to the single pump maximum. Higher fluctuation<br />

occurs at pressures below 100 psi.<br />

50.76 psi (3.5 bar)<br />

Temperature range 0 to 40°C<br />

Wetted materials in valve packages:<br />

Air valves<br />

Check valves<br />

Tubing and fittings<br />

The single pump maximum. Valves rated to 10,000 psi<br />

(689.5 bar)<br />

80 to 115 psi (5.5 to 7.9 bar)<br />

A1000 265.2<br />

A500 132.6<br />

A260 69.55<br />

A100X 32.50<br />

A100M 16.25<br />

A65DM 19.50<br />

A65 16.25<br />

45% of the single pump maximum rate. Cylinder cooling jackets<br />

should be used to obtain this rate.<br />

Hastelloy, PEEK, and PTFE<br />

SS316, sapphire, ruby, PEEK, PTFE<br />

SS316, gold<br />

8.3.1 Dual Air Valve<br />

Installation<br />

To install the air valve package (refer to Figure 8-3, 8-4, or 8-5,<br />

depending on your pump model):<br />

1. Position the valve bases 1.3 cm apart.<br />

2. Use the plugs to stopper the ports which will not be connected.<br />

DANGER<br />

RISK OF INJURY. THE PRESSURE PRODUCED<br />

COULD BE 700 BAR. PLEASE UTILIZE<br />

8-8