D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />
Section 8 Continuous Flow Introduction, Installation, & Operation<br />
8.3 Continuous Flow Air<br />
Valves<br />
Air Valve Package<br />
Numbers<br />
Air valves generally open and close faster and are less prone to<br />
error and component wear, making them suitable for industrial<br />
use and other applications where the system will be constantly<br />
running; however, they require a user-supplied pressurized air<br />
source of 80 to 115 psi (5.5 to 7.9 bar).<br />
A1000 68-1247-104<br />
A500 68-1247-061<br />
A65DM, A260, A100M, A100X 68-1247-058<br />
A65 68-1247-129<br />
Table 8-4 Continuous Flow Technical Specifications-Air Valves:<br />
Pressure fluctuation at switchover<br />
Minimum<br />
Maximum system backpressure<br />
Air supply source pressure<br />
Maximum flow rate (ml/min)<br />
Liquids: 65% of the single pump<br />
maximum rate.<br />
Liquefied gases<br />
5 psi (0.35 bar), at system backpressures from 100 psi<br />
(6.9 bar) to the single pump maximum. Higher fluctuation<br />
occurs at pressures below 100 psi.<br />
50.76 psi (3.5 bar)<br />
Temperature range 0 to 40°C<br />
Wetted materials in valve packages:<br />
Air valves<br />
Check valves<br />
Tubing and fittings<br />
The single pump maximum. Valves rated to 10,000 psi<br />
(689.5 bar)<br />
80 to 115 psi (5.5 to 7.9 bar)<br />
A1000 265.2<br />
A500 132.6<br />
A260 69.55<br />
A100X 32.50<br />
A100M 16.25<br />
A65DM 19.50<br />
A65 16.25<br />
45% of the single pump maximum rate. Cylinder cooling jackets<br />
should be used to obtain this rate.<br />
Hastelloy, PEEK, and PTFE<br />
SS316, sapphire, ruby, PEEK, PTFE<br />
SS316, gold<br />
8.3.1 Dual Air Valve<br />
Installation<br />
To install the air valve package (refer to Figure 8-3, 8-4, or 8-5,<br />
depending on your pump model):<br />
1. Position the valve bases 1.3 cm apart.<br />
2. Use the plugs to stopper the ports which will not be connected.<br />
DANGER<br />
RISK OF INJURY. THE PRESSURE PRODUCED<br />
COULD BE 700 BAR. PLEASE UTILIZE<br />
8-8