Section 8 Continuous Flow Introduction, Installation, & Operation<br />

8.2.2 Dual Check Valve<br />

Installation<br />

Check valve installation<br />

(refer to Figure 8-1 or 8-2, depending on your pump model):<br />

Note<br />

Tighten fittings just enough to hold tubing in place. Final tightening<br />

will be done after all the tubing and fittings are in place.<br />

1. Move the pumps so that the pump bases are about 1.3 cm<br />

apart.<br />

2. Use the plugs to stopper the ports which will not be connected<br />

to the continuous flow package.<br />

DANGER<br />

RISK OF INJURY. THE PRESSURE PRODUCED<br />

COULD BE 700 BAR. PLEASE UTILIZE<br />

APPROPRIATE TUBING AND CONNECTIONS<br />

NOTED IN THE MANUAL.<br />

3. Attach the connectors appropriate to your pump to the 6.4<br />

cm unbent stainless steel tubing lengths and use the 1 8"<br />

nuts and ferrules to attach to the double check valve housings.<br />

4. Attach the inlet tubing assembly. The tubing you use<br />

depends upon whether or not your source is pressurized.<br />

a. Pressurized Reservoir:<br />

Connect the 1 8" 1.5 m stainless steel tubing to the<br />

bottom of the inlet tee, using high pressure valve fittings.<br />

Note<br />

The filter and PTFE tubing are not used.<br />

b. Nonpressurized Reservoir:<br />

Connect the 1 8" 1.5 m PTFE tubing to the bottom of<br />

each inlet check valve using the ferrule and nut with<br />

support spring provided in the package. Attach the tubing<br />

to the filter connector and then connect the filter.<br />

Note<br />

The stainless steel tubing is not used.<br />

5. Once the contents of the continuous flow package are<br />

totally assembled, attach the connectors appropriate for<br />

your pump on one side of the assembly to a port. Then, supporting<br />

the tee/valve assembly attach the other connector<br />

to the port of the second pump.<br />

6. Attach the outlet fittings to the top of the upper tee. Fittings<br />

for 1 8" tubing, as well as an optional reducing fitting<br />

8-3