D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 8 Continuous Flow Introduction,<br />

Installation, & Operation<br />

8.1 Introduction A D <strong>Series</strong> continuous flow pumping system consists of two<br />

syringe pumps and a valve accessory package regulated by one<br />

controller. This system allows you to continuously deliver your<br />

liquefied gas or liquid under constant flow or constant pressure<br />

mode.<br />

Continuous flow can be used in either of two modes: continuous<br />

constant pressure or continuous constant flow. Both of these<br />

modes assume positive displacement of the piston, with the<br />

exception of continuous Receive Mode (for detailed information<br />

about Receive Mode, refer to Technical Bulletin TB02, available<br />

on the Teledyne <strong>Isco</strong> web site).<br />

In any syringe pump continuous flow system there is a flow<br />

irregularity which occurs at the time of switchover from one<br />

pump module to the other. This flow irregularity can be measured<br />

as a pressure fluctuation. Teledyne <strong>Isco</strong> pressure fluctuation<br />

at switchover is ~0.35 bar (at system backpressures from<br />

6.9 bar to the single pump maximum).<br />

Before programming continuous flow, appropriate valves must be<br />

connected, the pumps must contain fluid, and there must be<br />

backpressure for operation.<br />

The following sections provide part numbers,<br />

specifications, and installation steps for:<br />

Continuous Flow Check Valves, on page 8-2<br />

Continuous Flow Air Valves, on page 8-8<br />

Continuous Flow Electric Valves, on page 8-13<br />

User Supplied Valves, on page 8-16<br />

Following these sections in the manual is information<br />

about:<br />

Operation, on page 8-17<br />

Operating Tips and Guidelines, on page 8-18<br />

Special Features, on page 8-19<br />

Accessories, on page 8-20<br />

8-1