07.03.2014 Views

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 7 Gradient Pumping for Pressure, Flow, and Concentration Modes<br />

A gradient pumping system includes<br />

high-pressure, two-way valves that connect the<br />

pump inlets to fluid reservoirs, and the pump<br />

outlets to the gradient mixer (dual pump<br />

system) or other apparatus (single pump<br />

system). Each refill kit and outlet valve<br />

package contains one two-way valve; the gradient<br />

package contains two. The kits contain<br />

all tubing and hardware necessary for valve<br />

installation.<br />

Two-Way<br />

Valve part<br />

#209-0098-05<br />

Following installation, the tubing connections must be tested for<br />

leaks before any program is run. If a leak is found, tighten the<br />

connection slightly. If the leak persists, swage the connection<br />

again with a new ferrule. Refer to Technical Bulletin TB05 Field<br />

Verification Procedures at<br />

http://www.isco.com/sp_applications/#techbulletins for leak test<br />

procedures.<br />

7-3

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!