D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 6 Serial Interface<br />

LGUL,F:xx,S:xx,I:xxxxx,F:xxxxx,D:xxxx.x,<br />

Step Number (01 – 99)<br />

File (01 – 99)<br />

Initial Pressure<br />

Range = 00010 – Maximum<br />

Units = psi<br />

Pressure Duration<br />

Range = 0000.1 – 9999.0 minutes<br />

Final Pressure<br />

Range = 00010 – Maximum<br />

Units = psi<br />

Figure 6-9 Upload Commands - Single pump pressure<br />

programming<br />

6.5.4 Range Command The RANGE command provides scaling information for the<br />

system parameters. For example,<br />

6R006RANGEA34<br />

may return a string similar to the one shown in Figure 6-10.<br />

R02CP= 10000,F=2500000,R=30.00000,V=102.9600,40<br />

Checksum<br />

Max Volume = 102.96<br />

Units = milliliters<br />

Max Refill Rate = 30.0 milliliters/minute<br />

Max Flow Rate = 25 milliliters/minute<br />

Maximum Pressure = 10,000 psi<br />

Figure 6-10 Range Serial Commands<br />

6-22