Section 5 Pump Maintenance & Repair<br />
5.11.7 Limit Sensor<br />
Calibration<br />
Once the new limit sensor assembly is installed, the sensors<br />
must be calibrated. Before calibrating the limit sensors, be sure<br />
the cylinder is unscrewed several turns. Follow steps 8-10 in the<br />
previous section (5.11.6). Use the following procedure to calibrate<br />
the limit sensors:<br />
1. Connect the pressure cable and power cord and turn the<br />
controller ON.<br />
2. Press REFILL and enter the maximum rate.<br />
3. Press STOP when the interrupter flag is approximately 1 4"<br />
from the full (lower) sensor.<br />
4. Press ‘A’ (FLOW RATE), and use the number key to set 10<br />
(ml/min). Press the ENTER key.<br />
5. Press REFILL.<br />
After the lower limit sensor is interrupted, the motor will<br />
stop.<br />
6. Press CONST FLOW.<br />
7. Press ‘A’ (FLOW RATE) and select MAX.<br />
8. Press RUN.<br />
9. Press STOP when the volume counter in the upper<br />
right-hand corner of the LCD reads 005.00 ml or less.<br />
10. Press ‘A’ (FLOW RATE) and enter 5 ml/min.<br />
11. Press RUN .<br />
12. Press STOP when the counter reads 0.30 or less.<br />
13. Press ‘A’ (FLOW RATE) and enter 1 ml/min.<br />
14. Press RUN.<br />
15. When the counter counts from 000.01 to 000.00 press<br />
STOP.<br />
16. Lower the upper limit sensor to the point where the flag<br />
just interrupts the sensor. (“CYLINDER EMPTY” will<br />
flash on the screen.)<br />
17. Tighten the top upper limit mounting screw.<br />
18. Turn the controller to STANDBY and disconnect the pressure<br />
cable.<br />
19. Screw the cylinder into the cylinder mounting block until<br />
the cylinder snugly bottoms against the piston (the cylinder<br />
will no longer turn).<br />
20. Unscrew the cylinder a minimum of 1 2 turn, then line up<br />
the inlet and outlet ports as you had them before.<br />
21. Lock the cylinder by tightening the locking screw.<br />
22. Connect the pressure cable and power cord and turn the<br />
controller ON.<br />
23. Press the REFILL key. Wait until you have access to the<br />
second upper limit sensor mounting screw and tighten it.<br />
24. Install the covers.<br />
5-22