D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 5 Pump Maintenance & Repair<br />

5.11.7 Limit Sensor<br />

Calibration<br />

Once the new limit sensor assembly is installed, the sensors<br />

must be calibrated. Before calibrating the limit sensors, be sure<br />

the cylinder is unscrewed several turns. Follow steps 8-10 in the<br />

previous section (5.11.6). Use the following procedure to calibrate<br />

the limit sensors:<br />

1. Connect the pressure cable and power cord and turn the<br />

controller ON.<br />

2. Press REFILL and enter the maximum rate.<br />

3. Press STOP when the interrupter flag is approximately 1 4"<br />

from the full (lower) sensor.<br />

4. Press ‘A’ (FLOW RATE), and use the number key to set 10<br />

(ml/min). Press the ENTER key.<br />

5. Press REFILL.<br />

After the lower limit sensor is interrupted, the motor will<br />

stop.<br />

6. Press CONST FLOW.<br />

7. Press ‘A’ (FLOW RATE) and select MAX.<br />

8. Press RUN.<br />

9. Press STOP when the volume counter in the upper<br />

right-hand corner of the LCD reads 005.00 ml or less.<br />

10. Press ‘A’ (FLOW RATE) and enter 5 ml/min.<br />

11. Press RUN .<br />

12. Press STOP when the counter reads 0.30 or less.<br />

13. Press ‘A’ (FLOW RATE) and enter 1 ml/min.<br />

14. Press RUN.<br />

15. When the counter counts from 000.01 to 000.00 press<br />

STOP.<br />

16. Lower the upper limit sensor to the point where the flag<br />

just interrupts the sensor. (“CYLINDER EMPTY” will<br />

flash on the screen.)<br />

17. Tighten the top upper limit mounting screw.<br />

18. Turn the controller to STANDBY and disconnect the pressure<br />

cable.<br />

19. Screw the cylinder into the cylinder mounting block until<br />

the cylinder snugly bottoms against the piston (the cylinder<br />

will no longer turn).<br />

20. Unscrew the cylinder a minimum of 1 2 turn, then line up<br />

the inlet and outlet ports as you had them before.<br />

21. Lock the cylinder by tightening the locking screw.<br />

22. Connect the pressure cable and power cord and turn the<br />

controller ON.<br />

23. Press the REFILL key. Wait until you have access to the<br />

second upper limit sensor mounting screw and tighten it.<br />

24. Install the covers.<br />

5-22