D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 5 Pump Maintenance & Repair<br />

Figure 5-11 Installing the plastic mounts<br />

8. Be sure that the cylinder is unscrewed several turns. If you<br />

need to use a wrench, we recommend a strap wrench or<br />

wrenches in the Teledyne <strong>Isco</strong> wrench package (P/N<br />

68-1247-067 for 65DM, 100DM, 100DX, 260D, and 65D;<br />

P/N 68-1247-068 for 500D; P/N 60-1247-093 for 1000D)<br />

which will not mar the cylinder’s outer surface.<br />

9. Center the pump bottom limit sensor with respect to the<br />

mounting screws and tighten them.<br />

10. Raise the top limit sensor to the extent of the slots and<br />

tighten the upper screw.<br />

Now the new limit sensors MUST be calibrated, as explained in<br />

the following section.<br />