D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 5 Pump Maintenance & Repair<br />

If the conduit is fastened directly to the case, then it must<br />

be removed and the plastic mounting tabs included with<br />

the limit sensor assembly package should be installed<br />

using the existing screws and screw holes.<br />

To install the mounting tabs<br />

a. Remove the screws and the conduit.<br />

b. Then use the same screws and holes to install the<br />

mounts. Be sure the tab portion of the mount is facing<br />

away from the sensors, as shown in Figure 5-11, (the<br />

screw aperture is closest to the sensors).<br />

If your pump already has the plastic tabs installed, cut and<br />

discard the plastic cable ties which run through the tabs<br />

and hold the conduit.<br />

6. Remove the four screws holding the limit sensors, and<br />

unplug the cable from the power circuit board. (The<br />

replacement limit sensor assembly includes both the upper<br />

and lower sensor and the plug.)<br />

7. Install the new limit sensor harness; and secure the protective<br />

conduit by running the cable ties through the mounts<br />

and tightening. Then cut the excess cable tie and discard.<br />

Note<br />

Be sure the circuit board connector is mated pin-for-pin with<br />

the jack.<br />

Figure 5-10 Limit sensor replacement<br />

5-20