D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 5 Pump Maintenance & Repair<br />

3. Remove the four screws holding the pump base cover in<br />

place, and remove the cover to expose the PreAmp CBA<br />

(Figure 5-8, below).<br />

R6<br />

R27<br />

Figure 5-8 65D PreAmp CBA<br />

4. Connect the voltmeter to J14 (left red banana plug on the<br />

pump rear panel, Figure 5-9), and J16 (black center<br />

banana plug).<br />

J16<br />

J14<br />

Figure 5-9 Pressure output jacks, pump rear panel<br />

5. On the PreAmp CBA, adjust R6 to 0 VDC ±.010 at J14<br />

with reference to J16.<br />

6. On the controller, press ZERO PRESS, and zero the pressure<br />

for the appropriate pump.<br />

7. Enter a set pressure (such as 5,000 psi) and run the pump<br />

to that pressure. On the PreAmp CBA, adjust R27 until<br />

the pressure gauge reads the set pressure.<br />

8. Run the pump up to 10,000 psi, checking the pressure<br />

gauge for accuracy. If the reading is not within ±25 psi,<br />

readjust R27 to match the set pressure.<br />

5-18