07.03.2014 Views

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

D-Series Syringe Pumps - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 5 Pump Maintenance & Repair<br />

Note<br />

Parts for Configuration 1 and Configuration 2 are not interchangeable.<br />

5.10 Troubleshooting The following sections contain tables listing the most commonly<br />

used test points and their voltages. Refer to the controller schematic<br />

and the pump schematic (available online on our Web site).<br />

DANGER<br />

RISK OF ELECTRIC SHOCK -<br />

DISCONNECT THE ELECTRIC POWER BEFORE<br />

SERVICING. ONLY TRAINED SERVICE<br />

PERSONNEL MAY REMOVE THE CASE TOP.<br />

DANGER<br />

RISQUE DE CHOC ÉLECTRIQUE.<br />

COUPER L´ALIMENTATION AVANT LA<br />

RÉPARATION. L´USAGER NE DOIT PAS<br />

DÉMONTER L´INSTRUMENT OU DÉRANGER LE<br />

MÉCANISME DEDANS. ADRESSER LA<br />

REPARATION SEULEMENT AUX TECHNICIENS<br />

COMPÉTENTS.<br />

5.10.1 Power Supply (A1) Output voltages of the power supply can be verified. The voltages<br />

in the pump unit, between circuit common (TP106) and the fuses<br />

or at the test points should be:<br />

F104 22 VDC to 31 VDC<br />

F103 8.1 VDC to 13 VDC<br />

F102 22 VDC to 31 VDC<br />

TP111 15 VDC 0.6 VDC<br />

TP110 15 VDC 0.6 VDC<br />

TP109 5 VDC 0.3 VDC<br />

The voltage between motor common (TP101) and +VM (TP105)<br />

should be + 30 V to + 45 V (low speed) or +60 to +90 V (high<br />

speed). See section 4.<br />

5.10.2 Controller The voltages in the controller unit between circuit common<br />

(TP134) and the test points should be:<br />

TP106 5 VDC 0.002 VDC<br />

TP112 8.1 VDC to +13 VDC<br />

TP131 15 VDC 0.6 VDC<br />

TP135 15 VDC 0.6 VDC<br />

TP136 10 VDC to +12 VDC<br />

TP137 10 VDC to -12 VDC<br />

5-16