Section 5 Pump Maintenance & Repair<br />

Note<br />

It is very important that the shipment be well-packed and fully<br />

insured. Damage claims must be settled between you and the<br />

carrier. This can delay repair and return of the unit to you.<br />

5.1.3 Removing the Case Top For some maintenance procedures the case top of the pump component<br />

may need to be removed. Use the following procedures:<br />

DANGER<br />

RISK OF ELECTRIC SHOCK - DISCONNECT THE<br />

ELECTRIC POWER BEFORE SERVICING. ONLY<br />

TRAINED SERVICE PERSONNEL MAY REMOVE<br />

THE CASE TOP.<br />

DANGER<br />

RISQUE DE CHOC ÉLECTRIQUE. COUPER<br />

L´ALIMENTATION AVANT LA RÉPARATION.<br />

L´USAGER NE DOIT PAS DÉMONTER<br />

L´INSTRUMENT OU DÉRANGER LE MÉCANISME<br />

DEDANS. ADRESSER LA REPARATION<br />

SEULEMENT AUX TECHNICIENS COMPÉTENTS.<br />

1. Locate and remove the four case top screws on the sides of<br />

the instrument, two on each side of the cabinet.<br />

2. Pull the cover straight up and off.<br />

5.2 Lubrication<br />

Note<br />

The pump is a precision engineered instrument which must be<br />

lubricated after two years or every 6,000 strokes (whichever<br />

comes first) to assure proper service life.<br />

The pump has an easy-to-access lube wheel that keeps the main<br />

gears lubricated during operation. See Figure 5-1, configuration<br />

1 or 2, depending on your pump motor type. For your convenience,<br />

a lubrication kit (containing Never-seez and ALMASOL<br />

609 lubricants) is included in your pump accessory package.<br />