D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 3 Single & Multiple Independent Pump Operation<br />

3.9.1 Compatibility The hardware for this option includes a rear panel mounted<br />

circuit board which is compatible with the D <strong>Series</strong> 68-1240-026<br />

controller only.<br />

3.9.2 Hardware Three 12-bit digital-to-analog outputs represent the pump flow<br />

rates scaled to the maximum flow rate limits similar to the analog<br />

inputs. The three jumper selectable output ranges are 0 to 5V, 0 to<br />

10V, and 5V to 5V. These jumpers are accessed by removing the<br />

controller case top. Four outputs are provided on the circuit board<br />

and each output can be set to a different range, see Figure 3-9.<br />

The controller is shipped with all jumpers set for the 0 to 10V<br />

range. If less than three pumps are connected to the controller,<br />

pump volume delivered is provided by the extra outputs. The<br />

combinations are shown in Table 3-2.<br />

The MAX FLOW setting accessed via the LIMITS key, sets the<br />

fullscale flow rate. (See Section 3.5.4). For example, if the voltage<br />

output range is 5 volts and the desired output scale is 1 Volt per<br />

10 ml/min, the MAX FLOW limit should be set to 50 ml/min. The<br />

MAX FLOW setting may not exceed the pump specification.<br />

The volume analog outputs are not adjustable. The full-scale<br />

range represents one pump stroke. These outputs represent<br />

volume delivered, with the maximum output at cylinder empty.<br />

3.9.3 Current Loop Output The pump controller can be configured to provide a 4–20 mA<br />

current loop output. Contact the factory for more information on<br />

this option.<br />

Figure 3-9 Output range selection jumpers<br />