D <strong>Series</strong> <strong>Syringe</strong> <strong>Pumps</strong><br />

Section 3 Single & Multiple Independent Pump Operation<br />

Note<br />

If you make an error, press the CLEAR ENTRY key to delete<br />

your last keystroke. Each time you depress the CLEAR ENTRY<br />

key, you will delete back a character.<br />

4. Press the ENTER key once the desired pressure is displayed.<br />

5. Press the RUN key to initiate pump operation.<br />

3.7 External Control The pump can be externally controlled for pressure or flow rate<br />

operation with an analog voltage or through the serial interface.<br />

Controlling the pump with an analog voltage is discussed in<br />

section 3.3.8.<br />

The serial interface allows you to control the pump operation<br />

from an IBM-PC or compatible computer which has an RS-232-C<br />

serial output. The serial interface accepts English command<br />

words from the computer, like constant pressure, refill, etc. For<br />

more information see Section 6, Serial Interface.<br />

3.8 Remote RUN/STOP The D series syringe pump RUN/STOP function can be externally<br />

controlled by a switch contact closure or TTL input. The<br />

input voltage is 5 volts and is internally pulled high (RUN). The<br />

input is level sensitive (must remain high for RUN or low for<br />

STOP) and must be high for normal operation of serial (RS-232)<br />

control.<br />

To use the remote RUN/STOP feature, first press RUN or force<br />

the RUN/STOP pin low to enable the pump. Thereafter the<br />

RUN/STOP pin will control operation. Pressing STOP on the<br />

front panel will override the RUN/STOP pin.<br />

3.8.1 Wire Connections Two wires are required for external RUN/STOP control. The<br />

digital common or ground wire should be connected to one of the<br />

four DIGITAL GROUND terminals of the ACCESSORY connector<br />

on the controller rear panel. The control wire should be<br />

connected to terminal 1, under DIGITAL INPUT. If an electrically<br />

isolated relay is used, one relay terminal should be connected<br />

to digital ground and the other to terminal 1, under<br />

DIGITAL INPUT.<br />

If two pumps are used with the controller, the second control wire<br />

should be connected to terminal 2, under DIGITAL INPUT.<br />

If three pumps are used with the controller, the third control wire<br />

should be connected to terminal 3, under DIGITAL INPUT or to<br />

auxiliary (P11) pin 17.<br />

3.9 Analog Flow Rate and<br />

Volume Output Option<br />

The purpose of the analog output option (P/N 68-1247-070) is to<br />

provide for analog monitoring of the syringe pump flow rate. If<br />

two or less pump modules are connected to the controller, pump<br />

volume delivered can also be monitored via the analog output.<br />

These outputs are often used with analog based plant or process<br />

monitoring equipment.<br />