<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 5 Maintenance<br />

5.4.1 Replacing the<br />

Desiccant: <strong>AV</strong> <strong>Module</strong><br />

Collar<br />

The desiccant is contained in a cartridge located on the left side<br />

of the <strong>2150EX</strong>. To remove the cartridge, unscrew the collar and<br />

slide the cartridge out of the <strong>2150EX</strong>. The opaque tube reveals<br />

the silica gel desiccant inside.<br />

CAUTION<br />

To prevent static electricity, do not replace silica in potentially<br />

explosive atmospheres. Empty and fill the desiccant cartridge<br />

in a safe area.<br />

Cartridge<br />

5.4.2 Replacing the<br />

Desiccant: Battery<br />

<strong>Module</strong><br />

5.4.3 Reactivating the<br />

Desiccant<br />

To replace the silica gel desiccant:<br />

1. Hold the cartridge upright with the collar at the top.<br />

2. As shown to the left, push the collar off the cartridge.<br />

3. Empty the saturated silica gel beads or granules.<br />

4. Fill the tube with new (P/N 099-0011-03) or reactivated<br />

(see section 5.4.3) silica gel desiccant.<br />

5. Press the collar onto the tube.<br />

6. Slide the cartridge into the <strong>2150EX</strong> <strong>Module</strong>. Tighten the<br />

collar to seal the cartridge in place.<br />

A bag of desiccant is located inside each of the battery caps<br />

behind a retaining plate. To replace the desiccant:<br />

1. Loosen the two mounting screws that secure the metal<br />

retaining plate.<br />

2. Rotate the retaining plate until it is free from the mounting<br />

screws.<br />

3. Remove the spent desiccant bag from the cap and replace it<br />

with a new (P/N 099-0002-33) or reactivated (see section<br />

5.4.3) bag.<br />

4. Replace the retaining plate and secure it with the screws.<br />

Silica gel beads, granules, and bags of desiccant can be reactivated.<br />

CAUTION<br />

Desiccant may produce irritating fumes when heated. Observe<br />

the following precautions:<br />

• Use a vented oven in a well-ventilated room.<br />

• Do not remain in the room while the regeneration is taking<br />

place.<br />

• Use the recommended temperature. Avoid heating the<br />

desiccant at higher than recommended temperatures.<br />

5-9