2150EX AV Flow Module - Isco
2150EX AV Flow Module - Isco
2150EX AV Flow Module - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />
Section 5 Maintenance<br />
4. Replace the door. Align the small triangle on the door with<br />
the triangle above the battery port, push inward, and<br />
rotate 1 /4 turn clockwise so the curved arrow is at the top of<br />
the door.<br />
Repeat steps 1 through 4 to install the second battery pack.<br />
Note<br />
The battery packs should always be replaced as a pair. Never<br />
mix old and new batteries. Battery packs should be recycled as<br />
any normal lead-acid battery.<br />
Note<br />
For storage, the battery packs may be rotated 180 degrees<br />
and inserted into the 2191EX. This disconnects the packs for<br />
storage.<br />
5.4 Desiccant<br />
20 30<br />
40<br />
Battery <strong>Module</strong><br />
Humidity Indicator<br />
The <strong>2150EX</strong> System devices use desiccant to protect the internal<br />
components from moisture damage. In the <strong>2150EX</strong>, a desiccant<br />
cartridge is used to dry the reference air for the sensor. This prevents<br />
moisture from plugging the reference line, which would<br />
cause the sensor to report erroneous level readings. The cartridge<br />
is filled with indicating silica gel, which is blue or yellow<br />
when dry. As the desiccant becomes saturated, the color changes<br />
from blue to pink, or from yellow to green. Replace the desiccant<br />
before the entire length of the cartridge turns pink or green.<br />
The 2191EX battery module uses desiccant bags to keep the<br />
interior of the case dry. The bags are located inside the battery<br />
caps. Attached to the inside face of each cap is a humidity indicator.<br />
Humidity indicators have regions that display 20, 30, and<br />
40 percent humidity levels. Ideally, each region should be completely<br />
blue. As the desiccant becomes saturated, the humidity<br />
levels will increase and the regions turn pink. When the 40<br />
percent region begins to turn pink, the components are no longer<br />
adequately protected and the desiccant must be replaced.<br />
5-8