07.03.2014 Views

2150EX AV Flow Module - Isco

2150EX AV Flow Module - Isco

2150EX AV Flow Module - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 5 Maintenance<br />

4. Replace the door. Align the small triangle on the door with<br />

the triangle above the battery port, push inward, and<br />

rotate 1 /4 turn clockwise so the curved arrow is at the top of<br />

the door.<br />

Repeat steps 1 through 4 to install the second battery pack.<br />

Note<br />

The battery packs should always be replaced as a pair. Never<br />

mix old and new batteries. Battery packs should be recycled as<br />

any normal lead-acid battery.<br />

Note<br />

For storage, the battery packs may be rotated 180 degrees<br />

and inserted into the 2191EX. This disconnects the packs for<br />

storage.<br />

5.4 Desiccant<br />

20 30<br />

40<br />

Battery <strong>Module</strong><br />

Humidity Indicator<br />

The <strong>2150EX</strong> System devices use desiccant to protect the internal<br />

components from moisture damage. In the <strong>2150EX</strong>, a desiccant<br />

cartridge is used to dry the reference air for the sensor. This prevents<br />

moisture from plugging the reference line, which would<br />

cause the sensor to report erroneous level readings. The cartridge<br />

is filled with indicating silica gel, which is blue or yellow<br />

when dry. As the desiccant becomes saturated, the color changes<br />

from blue to pink, or from yellow to green. Replace the desiccant<br />

before the entire length of the cartridge turns pink or green.<br />

The 2191EX battery module uses desiccant bags to keep the<br />

interior of the case dry. The bags are located inside the battery<br />

caps. Attached to the inside face of each cap is a humidity indicator.<br />

Humidity indicators have regions that display 20, 30, and<br />

40 percent humidity levels. Ideally, each region should be completely<br />

blue. As the desiccant becomes saturated, the humidity<br />

levels will increase and the regions turn pink. When the 40<br />

percent region begins to turn pink, the components are no longer<br />

adequately protected and the desiccant must be replaced.<br />

5-8

Hooray! Your file is uploaded and ready to be published.

Saved successfully!