2150EX AV Flow Module - Isco
2150EX AV Flow Module - Isco
2150EX AV Flow Module - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />
Section 5 Maintenance<br />
2. Align the connectors and insert the new battery pack into<br />
the Battery <strong>Module</strong>.<br />
3. Check the humidity indicator disk inside the door. (See section<br />
5.4.2.)<br />
4. Replace the door. Align the small triangle on the door with<br />
the triangle above the battery port, push inward, and<br />
rotate 1 /4 turn clockwise so the curved arrow is at the top of<br />
the door.<br />
Repeat steps 1 through 4 to install the second battery pack.<br />
When finished, reconnect the <strong>2150EX</strong> module(s).<br />
Note<br />
The battery packs should always be replaced as a pair. Never<br />
mix old and new batteries. Battery packs should be disposed<br />
of according to local battery disposal regulations.<br />
Note<br />
For storage, the battery packs may be rotated 180 degrees<br />
and inserted into the 2191EX. This disconnects the packs for<br />
storage.<br />
5.3.2 SLA2191EX Lead-Acid<br />
Batteries<br />
The lead-acid battery packs should give you several weeks of<br />
service before they need recharged, depending upon your data<br />
storage intervals (see Table 1-5 and Figure 5-2).<br />
Figure 5-2 SLA2191EX Battery Pack Voltage Chart<br />
5-4