07.03.2014 Views

2150EX AV Flow Module - Isco

2150EX AV Flow Module - Isco

2150EX AV Flow Module - Isco

SHOW MORE
SHOW LESS

You also want an ePaper? Increase the reach of your titles

YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.

<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 5 Maintenance<br />

Instead, measure the voltage by disconnecting the <strong>2150EX</strong><br />

module(s) from the 2191EX battery module and attaching a volt<br />

meter to the connector on top of the 2191EX. Check the battery<br />

packs individually, not as a pair.<br />

WARNING<br />

When using the above method to check voltage, use care<br />

not to short any pins.<br />

Battery Replacement<br />

Fuse Replacement<br />

Batteries should be replaced according to the instructions below.<br />

The batteries should be replaced with:<br />

• two new lithium battery packs<br />

(P/N 68-2000-022).<br />

The fuses should be replaced only with:<br />

• Littel fuse #216.315<br />

(P/N 411-0922-30)<br />

WARNING<br />

Substitution of components may impair intrinsic safety.<br />

WARNING<br />

To avoid overloading the fuses in the lithium LTC2191EX<br />

battery packs, disconnect the <strong>2150EX</strong> module(s) before<br />

installing or replacing battery packs.<br />

Battery Pack<br />

Door<br />

Figure 5-1 Illustration of LTC2191EX Battery Packs<br />

To install the lithium LTC2191EX battery packs, first disconnect<br />

the <strong>2150EX</strong> module(s) and then:<br />

1. Remove the battery door. To remove the door, turn it 1 /4 turn<br />

counter-clockwise and pull it from the Battery <strong>Module</strong>.<br />

5-3

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!