<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 5 Maintenance<br />

5.1 Maintenance<br />

Overview<br />

This section explains the maintenance requirements of the<br />

<strong>2150EX</strong> Area Velocity <strong>Flow</strong> <strong>Module</strong>, 2191EX Battery <strong>Module</strong>,<br />

and the <strong>AV</strong><strong>2150EX</strong> Sensor.<br />

The <strong>2150EX</strong> System is designed to perform reliably in adverse<br />

conditions with a minimal amount of routine service requirements.<br />

To keep your system working properly, the following<br />

should be checked at regular intervals:<br />

• Battery power (section 5.3)<br />

• Desiccant (section 5.4)<br />

• Channel conditions (section 5.5)<br />

Maintenance intervals are affected by many variables; for<br />

example, the Data Storage Rate will affect the battery life.<br />

Humidity levels obviously affect the service life of the desiccant,<br />

and the amount of debris in the stream can drastically alter the<br />

channel conditions.<br />

As a guide, a basic system installed in an environment with moderate<br />

humidity levels and an <strong>AV</strong> Sensor installed in a channel<br />

relatively free from debris and silt, the maintenance interval<br />

should not exceed three months. A basic system is defined as:<br />

• a <strong>2150EX</strong> <strong>Module</strong> and <strong>AV</strong> Sensor,<br />

• a 2191EX Battery <strong>Module</strong> (including a fresh pair of<br />

removable LTC2191EX 8 volt lithium or SLA2191EX 8<br />

volt lead-acid battery packs) or a 2196EX Zone 1 Battery<br />

<strong>Module</strong> (includes two permanently installed<br />

rechargeable batteries)<br />

• recording readings at the default intervals of 15<br />

minutes.<br />

Experience is often the best tool to use when establishing<br />

minimum maintenance intervals for your system. Until you have<br />

gained an understanding of the <strong>2150EX</strong> <strong>Module</strong>’s operation<br />

under differing environmental conditions, a weekly maintenance<br />

interval is recommended.<br />