<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 3 Programming<br />

If the selected flow conversion requires channel dimensions,<br />

actual channel measurements should be taken. Channel measurements<br />

are preferred over nominal values. Significant errors may<br />

be introduced if your measurements are inaccurate. The example<br />

below illustrates the importance of accurate measurements.<br />

Example:<br />

Nominal Pipe Diameter: 10 inches<br />

Actual Pipe Diameter: 10.25 inches<br />

Level Measured Near Outfall: 2.75 inches<br />

Correct Level Measurement: 3 inches<br />

During programming, you enter 10 inches for the round pipe<br />

diameter - from the pipe manufacturer’s specification. You<br />

also enter the 2.75 inch level measurement taken behind the<br />

sensor near an outfall. Although each setting has only a 0.25<br />

inch error, the cumulative flow measurement error may<br />

exceed 14%!<br />

Refer to the information in Section 3.3.1 to determine where to<br />

measure the channel dimensions.<br />

3.3.5 Silt Level Silting in the flow stream will alter your channel dimensions,<br />

affecting the flow rate conversion. To compensate for a buildup of<br />

silt, a Silt Level value can be entered on the <strong>Flow</strong> Rate measurement<br />

tab in <strong>Flow</strong>link. Silt level compensation is only<br />

available when using Area Velocity flow conversion.<br />

3.3.6 Data Storage Rates The data storage function of a <strong>2150EX</strong> can record level, velocity,<br />

flow rate, total flow, and input voltage readings. The interval at<br />

which the <strong>2150EX</strong> stores the readings is called the Data Storage<br />

Rate. The <strong>2150EX</strong> is shipped with default storage rates of 15<br />

minutes for the level, velocity, and flow rate, and 1 hour for total<br />

flow and input voltage readings.<br />

You can modify the data storage rates to log readings at a faster<br />

or slower rate. Keep in mind that although the <strong>2150EX</strong> can store<br />

data as fast as 1 reading every 15 seconds, faster storage rates<br />

will shorten battery life, increase memory usage, and lengthen<br />

Retrieve Data (interrogation) times.<br />

You can also create conditional data storage rates. The <strong>2150EX</strong><br />

can log data at a secondary rate when user-defined conditions<br />

have been met. For example, a <strong>2150EX</strong> can store level readings<br />

at a primary rate of 15 minutes, and a secondary rate of 1 minute<br />

when the level reading is greater than or equal to 1 foot. Secondary<br />

rates allow you to collect detailed data when defined<br />

events of interest occur, while reducing power and memory consumption<br />

when detailed readings are not needed.<br />

To modify the Data Storage Rates, first click on the Set Up Data<br />

Storage… button on a measurement tab. Then enter the Primary<br />

and Secondary Rate settings on the Data Storage Setup window.<br />

Repeat this for each measurement type.<br />