<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 2 Preparation and Installation<br />

Note<br />

The <strong>2150EX</strong> requires <strong>Flow</strong>link 4.1 or later. If you require two<br />

minute data storage intervals, you will need version 4.16 or<br />

later.<br />

Refer to Section 3 and define the following properties:<br />

• Level - Enter a liquid level measurement to adjust the<br />

level readings from the <strong>AV</strong> Sensor.<br />

• Zero Level Offset – If the <strong>AV</strong> Sensor is not installed in<br />

the bottom-center of the channel, an offset distance<br />

must be entered.<br />

• Set <strong>Flow</strong> Rate to zero if no velocity data checkbox<br />

- Determines how the <strong>2150EX</strong> reports flow rates if<br />

stream velocity data is not available.<br />

• <strong>Flow</strong> Conversion – The <strong>2150EX</strong> can store flow rate<br />

readings. To correctly convert the measured level and<br />

velocity readings to a flow rate, the flow conversion<br />

method and channel properties should be defined.<br />

• Silt Level – (Area Velocity <strong>Flow</strong> Conversion Only) The<br />

<strong>2150EX</strong> can compensate for a build up of silt around the<br />

<strong>AV</strong> Sensor.<br />

These five settings should be considered a minimum<br />

requirement. Other settings, such as Data Storage Rates, Site<br />

Name, and <strong>Module</strong> Names, also may be set using <strong>Flow</strong>link.<br />