<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />
Section 2 Preparation and Installation<br />
CAUTION<br />
Under no circumstances should you leave any extra length of<br />
sensor cable dangling freely in the flow stream where it could<br />
trap debris or become tangled.<br />
Use gloves and eye protection when assembling and installing<br />
the rings in a pipe. Though deburred, the edges of the stainless<br />
steel can cut if improperly handled. Please read the information<br />
on how best to install this device.<br />
Observe general safety procedures when entering any manhole.<br />
See “General Safety Procedures” in the back of the manual<br />
for more information on general hazards and necessary<br />
precautions.<br />
2.9 Final Installation<br />
Check<br />
The system should be secured at the site. This prevents damage<br />
caused by accidental falls and from being swept away if the<br />
channel is flooded. In manholes, the module is often secured to a<br />
ladder rung. Teledyne <strong>Isco</strong>’s Customer Service Department or<br />
your local representative can assist you with installation options.<br />
As you complete the installation, the following should be checked<br />
before leaving the site unattended:<br />
1. The module should be positioned where it will be protected<br />
from submersion. Should the module become submerged,<br />
level readings may drift and the hydrophobic filter will seal<br />
to protect the reference air line.<br />
Note<br />
To protect the <strong>2150EX</strong> <strong>AV</strong> <strong>Flow</strong> module and sensor, the hydrophobic<br />
filter seals off the reference air line when it is exposed<br />
to excessive moisture. When sealed, the filter prevents irreparable<br />
damage, yet may cause the level readings to drift. This<br />
single-use filter must be replaced once it becomes sealed.<br />
2. Make sure all of the protective caps are in place. An<br />
unused upper communication connection must be capped<br />
to prevent damage and terminate the communication line.<br />
If the communication connector is in use, its cap should be<br />
properly stowed. Like the module and sensor connections,<br />
the protective caps and their O-rings should be cleaned<br />
and coated with a silicone lubricant. Damaged O-rings<br />
must be replaced (P/N 202-1006-69).<br />
3. Carefully route cables. Protect them from traffic in the<br />
area. Avoid leaving excess <strong>AV</strong> Sensor cable in the flow<br />
stream where it may collect debris.<br />
2.9.1 Program the <strong>Module</strong> After you have installed the <strong>AV</strong> Sensor in the flow stream, the<br />
flow stream properties must be defined. To do this, connect to the<br />
<strong>2150EX</strong> with <strong>Flow</strong>link software and define the stream properties<br />
in the <strong>2150EX</strong> module’s program settings. These ensure that the<br />
system correctly reads the liquid level and converts the measured<br />
level to flow rate.<br />
2-37