<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 2 Preparation and Installation<br />

CAUTION<br />

Under no circumstances should you leave any extra length of<br />

sensor cable dangling freely in the flow stream where it could<br />

trap debris or become tangled.<br />

Use gloves and eye protection when assembling and installing<br />

the rings in a pipe. Though deburred, the edges of the stainless<br />

steel can cut if improperly handled. Please read the information<br />

on how best to install this device.<br />

Observe general safety procedures when entering any manhole.<br />

See “General Safety Procedures” in the back of the manual<br />

for more information on general hazards and necessary<br />

precautions.<br />

2.9 Final Installation<br />

Check<br />

The system should be secured at the site. This prevents damage<br />

caused by accidental falls and from being swept away if the<br />

channel is flooded. In manholes, the module is often secured to a<br />

ladder rung. Teledyne <strong>Isco</strong>’s Customer Service Department or<br />

your local representative can assist you with installation options.<br />

As you complete the installation, the following should be checked<br />

before leaving the site unattended:<br />

1. The module should be positioned where it will be protected<br />

from submersion. Should the module become submerged,<br />

level readings may drift and the hydrophobic filter will seal<br />

to protect the reference air line.<br />

Note<br />

To protect the <strong>2150EX</strong> <strong>AV</strong> <strong>Flow</strong> module and sensor, the hydrophobic<br />

filter seals off the reference air line when it is exposed<br />

to excessive moisture. When sealed, the filter prevents irreparable<br />

damage, yet may cause the level readings to drift. This<br />

single-use filter must be replaced once it becomes sealed.<br />

2. Make sure all of the protective caps are in place. An<br />

unused upper communication connection must be capped<br />

to prevent damage and terminate the communication line.<br />

If the communication connector is in use, its cap should be<br />

properly stowed. Like the module and sensor connections,<br />

the protective caps and their O-rings should be cleaned<br />

and coated with a silicone lubricant. Damaged O-rings<br />

must be replaced (P/N 202-1006-69).<br />

3. Carefully route cables. Protect them from traffic in the<br />

area. Avoid leaving excess <strong>AV</strong> Sensor cable in the flow<br />

stream where it may collect debris.<br />

2.9.1 Program the <strong>Module</strong> After you have installed the <strong>AV</strong> Sensor in the flow stream, the<br />

flow stream properties must be defined. To do this, connect to the<br />

<strong>2150EX</strong> with <strong>Flow</strong>link software and define the stream properties<br />

in the <strong>2150EX</strong> module’s program settings. These ensure that the<br />

system correctly reads the liquid level and converts the measured<br />

level to flow rate.<br />

2-37