<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 2 Preparation and Installation<br />

The scissors mechanism has stainless steel ground terminals<br />

mounted on the hinged bracket for the attachment of bonding<br />

conductors.<br />

Figure 2-26 Scissors mechanism ground terminals<br />

2.8.3 Completing the <strong>AV</strong><br />

Sensor Installation<br />

The <strong>AV</strong> sensor installation is finished by securing any excess<br />

sensor cable using cable clamps or other means.<br />

The reference tube inside the cable can be restricted or blocked if<br />

the cable is kinked, sharply bent, coiled, or otherwise pinched.<br />

The sensor cable should be handled and mounted with care. Also,<br />

if there is any appreciable distance between the point where the<br />

sensor cable leaves the mounting apparatus and the location of<br />

the flow meter, be sure to attach the cable to the flow stream wall<br />

to prevent it from vibrating, moving around, tangling, or possibly<br />

collecting debris.<br />

WARNING<br />

Do not coil the sensor cable. This will form an inductor and<br />

create hazard.<br />