<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 2 Preparation and Installation<br />

WARNING<br />

The RS485EX cable maximum Um = 16.6V. If the RS232<br />

cable is used to connect a laptop computer to the Field<br />

Wizard or other <strong>Isco</strong> network device, NEVER use mains<br />

power to power the laptop; use battery power only.<br />

Note<br />

You can safely connect and disconnect the RS485EX cable<br />

from the EX Network cable without removing the <strong>2150EX</strong> module<br />

from the potentially explosive atmosphere.<br />

2.6.4 RS232 Communication<br />

Cable<br />

In a permanent installation, the 2194EX serves as the network<br />

communication isolator, with RS232 or RS485 communication<br />

via the top connector.<br />

Figure 2-21 RS232 communication cable<br />

The 9-pin connector of <strong>Isco</strong>’s RS232 communication cable<br />

(60-2004-046) connects to a computer’s serial port, and the<br />

push-on end connects to the top of the 2194EX.<br />