<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 2 Preparation and Installation<br />

Figure 2-17 Disassembling network cable connector<br />

2.6 Network<br />

Communication<br />

To connect the <strong>2150EX</strong> for network communication, one or two of<br />

three different cables are required, depending on the type of communication,<br />

whether the installation is portable or permanent,<br />

and whether or not the flow module installation is in a hazardous<br />

area. When the communication connector is not in use, it should<br />

always be capped to prevent corrosion and improve communications.<br />

When the communication connector is in use, store the cap<br />

on the holder next to the connector.<br />

Note<br />

Caps PUSH ON and PULL OFF. Do not rotate the caps to<br />

remove them from the connectors.<br />

2.6.1 EX Network Cable The EX Network cable (2m P/N 60-2004-335,<br />

8m P/N 60-2004-336) connects to the top of the <strong>2150EX</strong> stack and<br />

extends to the interface of the safe and hazardous areas, where<br />

the actual isolation is located.<br />