<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />
Section 2 Preparation and Installation<br />
Two interface cable assemblies are available from Teledyne <strong>Isco</strong>:<br />
The CA75EX-CTL, 75 meter (60-2004-337) provides the<br />
maximum length for powering two <strong>2150EX</strong> modules with<br />
sensors. The CA150EX-CTL, 150 meter (60-2004-338) provides<br />
the maximum length for powering one <strong>2150EX</strong> module with<br />
sensor. Maximum length may require reduced-power program<br />
settings. Contact the factory for custom-ordered lengths less than<br />
standard. Install the interface cables without coiling and cut to<br />
the appropriate practical length. Wire the interface cable to the<br />
2194EX using the Network Interface Connector Wiring Kit<br />
(60-2004-468), referring to Figures 2-13 through 2-17. Always<br />
comply with the clearly labeled entity parameters for hazardous<br />
location installations.<br />
Locking<br />
ring<br />
Socket<br />
insert<br />
Locking<br />
cap<br />
Main<br />
body<br />
Gland<br />
Gland<br />
cage<br />
Gland nut<br />
(When using conduit,<br />
replace with appropriate<br />
conduit fitting.)<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
<br />
Figure 2-14 Network cable connector and wiring<br />
Observe intrinsic safety requirements regarding proximity to<br />
external sources of potential electric or magnetic interference.<br />
Refer to IEC 10079-14 Section 12.2.2.5 on installation of cables<br />
and wiring.<br />
WARNING<br />
Do not coil the interface cable; this will form an inductor<br />
and create a hazard. The cable should be kept as short as<br />
is practical.<br />
Teledyne <strong>Isco</strong> strongly recommends that you route the<br />
interface cable through conduit between the safe and hazardous<br />
areas. Two different sizes of conduit fittings are provided<br />
with the connector wiring kit (Figure 2-15).<br />
2-22