<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 2 Preparation and Installation<br />

Two interface cable assemblies are available from Teledyne <strong>Isco</strong>:<br />

The CA75EX-CTL, 75 meter (60-2004-337) provides the<br />

maximum length for powering two <strong>2150EX</strong> modules with<br />

sensors. The CA150EX-CTL, 150 meter (60-2004-338) provides<br />

the maximum length for powering one <strong>2150EX</strong> module with<br />

sensor. Maximum length may require reduced-power program<br />

settings. Contact the factory for custom-ordered lengths less than<br />

standard. Install the interface cables without coiling and cut to<br />

the appropriate practical length. Wire the interface cable to the<br />

2194EX using the Network Interface Connector Wiring Kit<br />

(60-2004-468), referring to Figures 2-13 through 2-17. Always<br />

comply with the clearly labeled entity parameters for hazardous<br />

location installations.<br />

Locking<br />

ring<br />

Socket<br />

insert<br />

Locking<br />

cap<br />

Main<br />

body<br />

Gland<br />

Gland<br />

cage<br />

Gland nut<br />

(When using conduit,<br />

replace with appropriate<br />

conduit fitting.)<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

<br />

Figure 2-14 Network cable connector and wiring<br />

Observe intrinsic safety requirements regarding proximity to<br />

external sources of potential electric or magnetic interference.<br />

Refer to IEC 10079-14 Section 12.2.2.5 on installation of cables<br />

and wiring.<br />

WARNING<br />

Do not coil the interface cable; this will form an inductor<br />

and create a hazard. The cable should be kept as short as<br />

is practical.<br />

Teledyne <strong>Isco</strong> strongly recommends that you route the<br />

interface cable through conduit between the safe and hazardous<br />

areas. Two different sizes of conduit fittings are provided<br />

with the connector wiring kit (Figure 2-15).<br />