<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 2 Preparation and Installation<br />

2.5 Permanent<br />

Installations<br />

For permanent installations, the <strong>2150EX</strong> can be powered from a<br />

safe area by an associated apparatus, the 2194EX module. The<br />

2194EX also serves as a network interface, with network and<br />

RS232 communication via the top connector. The 2194EX<br />

requires 12 or 24 volts DC, and may be powered by an <strong>Isco</strong><br />

910/920 series power pack, 934 NiCad battery, or 940 series lead<br />

acid battery using power adapter cable 69-2004-451. For details<br />

about these power supplies, see <strong>Isco</strong>’s Power Products Guide<br />

(60-9003-092).<br />

Pin B = +12 Volts<br />

Pin A = Neutral<br />

Figure 2-12 Amphenol connector pins for 12V adapter cable 69-2004-451<br />

Note<br />

<strong>Isco</strong> AC power supplies do not provide galvanic isolation in<br />

accordance with IEC 60079-14 for Zone 0 installations.<br />

Grounding<br />

terminals<br />

EX Bottom Plate 60-2004-344<br />

Important Information<br />

Regarding "X" Marking<br />

Make sure the <strong>2150EX</strong> is secured so that it will not accidentally<br />

fall or be swept away by flooding. Mount the <strong>2150EX</strong> onto the EX<br />

bottom plate (60-2004-344) for suspension over the flow stream.<br />

Use the notched holes in the plate to insert fasteners to secure<br />

the module to a wall, or attach a carrying handle and suspension<br />

handle (60-2004-377 and 60-2004-386), which can be secured to a<br />

ladder rung. The EX bottom plate has stainless steel ground terminals<br />

for attaching bonding conductors in permanent installations<br />

where circulating current presents a hazard.<br />

The ATEX labeling on the serial tag of the 2194EX module shows<br />

a number ending in "X". The X marking indicates that there are<br />

special conditions that must be met to ensure intrinsic safety, as<br />

explained on page viii in the front of this manual.<br />

In the case of the 2194EX, this associated apparatus does not<br />

provide the galvanic isolation required for zone 0 installations in<br />

accordance with IEC 60079-14 (refer to IEC 60079-14 sections<br />

dealing with earthing of intrinsically safe circuits and installations<br />

for zone 0) when powered by an <strong>Isco</strong> AC power source.<br />