Section 2 Preparation and Installation<br />
–<br />
+<br />
2196EX port connector<br />
As the 2196EX output voltage decreases to a value near the<br />
7-volt shut-off threshold, <strong>Flow</strong>link software will issue a<br />
low-voltage warning (Figure 2-5).<br />
Neither of these protective functions should become necessary<br />
on a regular basis. Cycle life is improved by regularly<br />
re-charging the batteries before a drop in voltage is<br />
indicated.<br />
The voltage may also be tested with a voltmeter at the port connector<br />
on pins c (+) and d (–), or on the internal circuit board (see<br />
the test points shown in Figure 2-7).<br />
Note<br />
There is a 60K ohm resistor in series with the voltage sensing<br />
circuit. The voltage reading measured on the circuit board may<br />
have slight variance, depending on the voltmeter used.<br />
CAUTION<br />
A 2196EX module kept in storage for extended periods should<br />
be recharged approximately every six to nine months. The battery<br />
voltage should never be allowed to fall below 10.5 volts<br />
before recharging. Deep discharge of the lead-acid batteries<br />
can lead to permanent loss of capacity.<br />
WARNING<br />
Ensure that the 2196EX module case is never subjected to<br />
physical impact with enough force to cause cracking during<br />
transport, installation, operation, or storage. Damage to the<br />
case can compromise the unit’s safety.<br />
Important Information<br />
Regarding “X” Marking<br />
Read all labels carefully before installing the equipment!<br />
The <strong>2150EX</strong> and associated equipment are clearly labeled with<br />
color and/or text so you know what can be located in a safe or<br />
hazardous area (see figure below).<br />
The ATEX labeling on the serial tag of the 2196EX module shows<br />
a number ending in “X.” The X marking indicates that there are<br />
special conditions that must be met to ensure safety, as explained<br />
beginning on page viii in the front of this manual.<br />
In the case of the 2196EX, the module must be protected from<br />
mechanical damage when installed and during transit.<br />