2150EX AV Flow Module - Isco
2150EX AV Flow Module - Isco
2150EX AV Flow Module - Isco
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />
Section 2 Preparation and Installation<br />
Interrogator Cable<br />
2194EX Network/<br />
Power <strong>Module</strong><br />
Computer running<br />
<strong>Flow</strong>link<br />
(Enclosure)<br />
Power Adapter<br />
Cable<br />
(Hazardous boundaries are normally<br />
specified by local authorities.)<br />
SAFE AREA<br />
Mains (optional)<br />
POTENTIALLY EXPLOSIVE AREA<br />
<strong>Isco</strong> Power Pack<br />
CA75EX-CTL/<br />
CA150EX-CTL<br />
Interface Cable<br />
and Conduit<br />
<strong>2150EX</strong> Area Velocity<br />
<strong>Flow</strong> <strong>Module</strong><br />
with Bottom Plate<br />
This figure is not intended to<br />
depict the meeting of special<br />
conditions indicated by "X"<br />
markings on the equipment.<br />
Refer to IEC 60079-14 section<br />
12.2.4 regarding intrinsically safe<br />
apparatus that does not<br />
withstand the 500VAC electrical<br />
strength test.<br />
See the warnings below.<br />
Mounting Ring<br />
(See WARNINGS<br />
below.)<br />
FLOW<br />
<strong>AV</strong><strong>2150EX</strong> Sensor<br />
Figure 2-3 Typical Round-pipe Installation Connected to a 2194EX <strong>Module</strong> and Laptop<br />
(Permanent Installation, see Section 2.5)<br />
WARNING<br />
Due to the creation of a permanent grounding point between<br />
the sensor’s transducer cover and the mounting ring when<br />
the sensor is installed, the <strong>2150EX</strong> system can not<br />
withstand the 500 VAC test according to EN60079-11:2007<br />
clause 6.3.12. Refer to IEC 60079-14, section 12.2.4,<br />
regarding earthing of intrinsically safe circuits.<br />
WARNING<br />
The sensor mounting ring is a potential isolated charge<br />
carrier. Your installation MUST satisfy earthing<br />
requirements. Refer to IEC 60079-14 section 12.2.4 and IEC<br />
60079-11.<br />
2-7