<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />

Section 2 Preparation and Installation<br />

Interrogator Cable<br />

2194EX Network/<br />

Power <strong>Module</strong><br />

Computer running<br />

<strong>Flow</strong>link<br />

(Enclosure)<br />

Power Adapter<br />

Cable<br />

(Hazardous boundaries are normally<br />

specified by local authorities.)<br />

SAFE AREA<br />

Mains (optional)<br />

POTENTIALLY EXPLOSIVE AREA<br />

<strong>Isco</strong> Power Pack<br />

CA75EX-CTL/<br />

CA150EX-CTL<br />

Interface Cable<br />

and Conduit<br />

<strong>2150EX</strong> Area Velocity<br />

<strong>Flow</strong> <strong>Module</strong><br />

with Bottom Plate<br />

This figure is not intended to<br />

depict the meeting of special<br />

conditions indicated by "X"<br />

markings on the equipment.<br />

Refer to IEC 60079-14 section<br />

12.2.4 regarding intrinsically safe<br />

apparatus that does not<br />

withstand the 500VAC electrical<br />

strength test.<br />

See the warnings below.<br />

Mounting Ring<br />

(See WARNINGS<br />

below.)<br />

FLOW<br />

<strong>AV</strong><strong>2150EX</strong> Sensor<br />

Figure 2-3 Typical Round-pipe Installation Connected to a 2194EX <strong>Module</strong> and Laptop<br />

(Permanent Installation, see Section 2.5)<br />

WARNING<br />

Due to the creation of a permanent grounding point between<br />

the sensor’s transducer cover and the mounting ring when<br />

the sensor is installed, the <strong>2150EX</strong> system can not<br />

withstand the 500 VAC test according to EN60079-11:2007<br />

clause 6.3.12. Refer to IEC 60079-14, section 12.2.4,<br />

regarding earthing of intrinsically safe circuits.<br />

WARNING<br />

The sensor mounting ring is a potential isolated charge<br />

carrier. Your installation MUST satisfy earthing<br />

requirements. Refer to IEC 60079-14 section 12.2.4 and IEC<br />

60079-11.<br />

2-7