<strong>2150EX</strong> Area Velocity <strong>Flow</strong> System<br />
Section 2 Preparation and Installation<br />
2.2.2 Locating the Site The <strong>2150EX</strong> is designed to measure flow in open channels with<br />
or without a primary device. A primary device is a hydraulic<br />
structure, such as a weir or a flume that modifies a channel so<br />
there is a known relationship between the liquid level and the<br />
flow rate. Although the <strong>2150EX</strong> supports flow rate conversion in<br />
channels with a primary device, its level and velocity measurement<br />
capabilities are best suited for channels without a<br />
primary device.<br />
Note<br />
Primary devices limit the usefulness of the <strong>AV</strong> Sensor’s readings.<br />
In most cases, levels and velocities near these structures<br />
do not represent what normally occurs in the channel. If you<br />
must use area velocity flow conversion, or if your interest is the<br />
stream’s velocity, do not install the <strong>AV</strong> Sensor near a primary<br />
device. Move the <strong>AV</strong> Sensor away to where the flow is unaffected<br />
by the primary device.<br />
2.2.3 Channels Without a<br />
Primary Device<br />
2.2.4 Channels With a<br />
Primary Device<br />
When the <strong>AV</strong> Sensor is installed without a primary device, find a<br />
section of channel with a minimum of disturbances to the flow.<br />
Avoid areas with elbows, outfalls, inverts, junctions, etc. that<br />
create turbulence near the <strong>AV</strong> Sensor. The <strong>AV</strong> Sensor should be<br />
located away from these disturbances to a point where the flow<br />
has stabilized. For best results, install the <strong>AV</strong> Sensor where the<br />
flow is most uniform. Uniform flow is a condition where the<br />
water surface is parallel to the bottom of the channel.<br />
If the <strong>AV</strong> Sensor is installed in a primary device, its location<br />
depends on the type of primary device. Most primary devices<br />
have a specific place for the head (level) measurement sensor. For<br />
more details about the location of the head measuring point,<br />
refer to the <strong>Isco</strong> Open Channel <strong>Flow</strong> Measurement Handbook, or<br />
to information provided by the manufacturer of the primary<br />
device.<br />
Note<br />
When you install the <strong>AV</strong> Sensor for use within a primary<br />
device, a Level-to-<strong>Flow</strong> conversion method should be used.<br />
(See Programming, Section 3.)<br />
2.2.5 <strong>2150EX</strong> and <strong>AV</strong> Sensor<br />
Mounting<br />
Considerations<br />
Ideal sites are easily accessible for service and data collection,<br />
while still providing protection for the <strong>2150EX</strong> module devices.<br />
The <strong>2150EX</strong> module devices are rated NEMA 4X, 6P, and constructed<br />
of materials that can withstand harsh environments.<br />
However, continual exposure to UV light, or periodic submersion<br />
should be avoided to extend the life of the components.<br />
Typically, the <strong>2150EX</strong> is suspended inside a manhole. Suspending<br />
the <strong>2150EX</strong> near the opening will protect it from the elements,<br />
minimize the chance of submersion, and allow it to be<br />
easily retrieved without entering the manhole.<br />
2-3