Site Window - 4100/4200/6700<br />
Set flow rate to zero<br />
if no velocity data<br />
Password<br />
Occasionally velocity readings are lost because either a flow<br />
stream does not contain enough reflective particles, or the sensor<br />
is covered with silt. These lost readings are logged as “No Data<br />
Code 248,” and the flow meter is unable to calculate the flow rate.<br />
This check box selects how flow rates are reported during this<br />
condition.<br />
Checked, the flow meter stores the flow rate as 0.0 under this<br />
condition.<br />
Unchecked, the flow meter uses the current level reading and<br />
repeats the last valid velocity reading to calculate the flow rate.<br />
If used, a numeric password is stored in the instrument’s<br />
memory and a copy is stored in the <strong>Flowlink</strong> site. While<br />
connecting to the instrument, the passwords are compared. A<br />
match will allow you to make changes to the instrument’s<br />
program settings.<br />
If the passwords do not match, you will still be able to retrieve<br />
data from the instrument. However, program settings cannot be<br />
unlocked until the instrument is accessed by the correct site, or<br />
until the password is entered.<br />
Passwords protect the program settings <strong>Flowlink</strong> sends to flow<br />
meters. The password prevents you or others from accidentally<br />
using the wrong site to connect with the flow meter, and it<br />
protects the settings from being changed by an unauthorized<br />
party.<br />
The Change Password button on the Site Setup tab opens the<br />
Change Password window. From this window you can enable<br />
password protection and change an existing password.<br />
To enable password protection:<br />
1. Enter a numerical password in the “new password” field.<br />
2. Confirm your password by retyping it in the verify field.<br />
3. Press the OK button. <strong>Flowlink</strong> writes the password to the<br />
instrument and enables password protection.<br />
To change an existing password:<br />
1. Enter the existing password in the “old password” field.<br />
2. Enter the new password in the “new password” field.<br />
3. Confirm your new password by retyping it in the verify<br />
field.<br />
4. Press the OK button. <strong>Flowlink</strong> writes the new password<br />
to the instrument. The instrument will remain enabled.<br />
The Remove Password button on the Site Setup tab opens the<br />
Remove Password window.<br />
To remove password protection:<br />
1. Enter the password in the “old password” field.<br />
2. Press the OK button. <strong>Flowlink</strong> clears the password from<br />
the instrument and disables password protection.<br />