<strong>Isco</strong> <strong>Flowlink</strong> <strong>5.1</strong> ® <strong>Software</strong><br />

Site Window - 4100/4200/6700<br />

• Variable Blanking is only applicable to Ultrasonic<br />

instruments. Enter the variable blanking distance.<br />

Minimum depth for velocity Box - displays Minimum depth<br />

for velocity. This setting is only applicable for area velocity<br />

instruments with a standard or extended range sensor.<br />

Set flow rate to zero if no velocity data Check box - controls<br />

what an area velocity flow meter stores as flow rate readings<br />

when it is unable to measure velocity.<br />

View last Box - Contains three buttons, Report, History, and<br />

Program. These buttons are active if a text file was downloaded<br />

during the last data retrieval. This communication option must<br />

first be enabled.<br />

Password Box - Reports the current status of the instrument’s<br />

password protection and provides buttons to change and remove<br />

the password.<br />

Totalizers Box - Contains two buttons to reset the totalizers to<br />

zero.<br />

• Zero Total Flow button resets the cumulative flow<br />

reported in the Measurements tab of the Site Window.<br />

• Zero Sampler Enable Total Flow button resets the<br />

sampler enable flow count to zero.<br />

Press the Configure Like Another button to configure the<br />

current site so it has the same settings as another. (The site<br />

name will not be changed.) After pressing the button, you will be<br />

prompted to select a site from the Select A Site dialog box. Select<br />

the appropriate site and click OK.<br />

Variable Blanking<br />

Minimum depth for<br />

velocity<br />

Ultrasonic instruments can vary the blanking to avoid erroneous<br />

level readings. Proper selection of a value will ensure that the<br />

flow meter accepts only valid echoes from the surface of the<br />

liquid, while disregarding invalid returns from the walls, sides of<br />

the channel, manhole rungs, or other obstructions.<br />

Enter this value as the distance from the face of the ultrasonic<br />

level sensor to the maximum expected liquid level. Refer to the<br />

instrument’s instruction manual to calculate this distance. The<br />

flow meter will then ignore any echo reflected from this distance<br />

or less. Keep in mind that if the liquid level rises into the<br />

variable blanking distance, the flow meter will ignore the return<br />

from the liquid surface.<br />

For area velocity instruments with a standard or extended range<br />

sensor, this sets the minimum depth for measured velocity<br />

readings. Below this level, the flow meter approximates the<br />

velocity using earlier readings, and will not be able to detect the<br />

direction of flow. Refer to the instrument’s instruction manual to<br />

determine the correct setting.<br />

Area velocity flow meters using a low-profile sensor do not<br />

require a minimum depth setting. Low profile sensors always use<br />

one inch (25 mm).<br />

79