<strong>Isco</strong> <strong>Flowlink</strong> <strong>5.1</strong> ® <strong>Software</strong><br />

Site Window - 2100<br />

That is, the sampler would stay disabled until a<br />

significant rainfall event occurred. This Rain Event<br />

condition would read “When 2105 Interface<br />

Module:Rainfall is greater than or equal to 0.25 inches<br />

in a 60 minute period following a ‘dry’ period of 1 Week.”<br />

To define the Rain Event condition:<br />

1. Select a 2105 Rainfall measurement from the pull down<br />

list.<br />

2. Enter the threshold amount of rainfall and units of<br />

measure from the pull down list.<br />

3. Enter a time duration in minutes to define the rate of<br />

increased rainfall.<br />

4. Define the “dry” period (period when rate of rainfall is<br />

below the threshold).<br />

NOTE: To specify a Rain Event for any time rainfall is at or<br />

above the threshold, enter a “dry” period of zero.<br />

Report<br />

Network Resolution<br />

- 2100<br />

This window displays text reports downloaded from the<br />

instrument or a list of alarm events from an attached 2103 or<br />

2105 modem module. Refer to the instrument’s instruction<br />

manual for more information on reading the individual reports.<br />

For instrument reports, this window has a 50,000 character limit<br />

and may cause very large reports or histories to appear<br />

incomplete. To view complete text reports exceeding this limit, or<br />

to print the report, click the Save To File button. Then, open the<br />

file in a text editor such as Notepad, or your word processing<br />

application.<br />

For alarm-event lists, there is a 20 line alarm-event limitation.<br />

After the 20 alarm-event limit is reached, the latest alarm event<br />

will be posted and the oldest alarm event will be removed from<br />

the list.<br />

The Network Resolution screen will appear when a 2100 module<br />

has been replaced or a 2100 module has been added to the<br />

network. You need to indicate what action should be taken to<br />

resolve the situation.<br />

Connected Modules - The physical module(s) that were found<br />

at the site are listed on the left side of the screen. If you click the<br />

ID button to the left of a listed module, the module will be<br />

identified by illuminating its LED or performing an identifying<br />

action (such as a beep from the Field Wizard).<br />

Database Site Definitions - The modules that <strong>Flowlink</strong> knows<br />

were previously at the site are listed on the right side of the<br />

screen. Where possible, these will be matched with the connected<br />

modules listed on the left side of the screen.<br />

If you click the ! button to the right of module, a window will be<br />

displayed that contains information about the connected module<br />

and selected module from the database. This information<br />

includes the Site Name, Module Name, Module Type, Module<br />

72