Flowlink 5.1 Software Instruction Manual - Isco
Flowlink 5.1 Software Instruction Manual - Isco
Flowlink 5.1 Software Instruction Manual - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>Isco</strong> <strong>Flowlink</strong> <strong>5.1</strong> ® <strong>Software</strong><br />
Site Window - 2100<br />
Note: Contact your Server Administrator for the Modem<br />
and IP listener information.<br />
After setting up the Pushed Data rates, click OK to accept the<br />
settings, and Apply (F9) the new settings to the module.<br />
Pushing Data from<br />
4200 and 6700<br />
Series Instruments<br />
(2105 only)<br />
Measurements -<br />
2100<br />
Level - 2100<br />
The 2105 module can capture, log, and push data from 4200<br />
Series flow meters and 6700 Series samplers. The 2105<br />
interfaces with the instrument through a simulated Modbus<br />
connection. 4200/6700 data parameters must be entered as<br />
Modbus parameters.<br />
For complete Modbus input information for 4200/6700<br />
instruments, see your 2105 instruction manual.<br />
This tab of the Site window displays the settings for the selected<br />
measurement type. The module and its functions determine<br />
which settings are available. This tab is variable, meaning its<br />
label and content will change according to which Module and<br />
measurement type you have selected.<br />
From this tab you may:<br />
• View the Module and Measurement name.<br />
• View the current reading.<br />
• Modify the measurement settings and unit of measure (if<br />
user-configurable).<br />
• Setup Data Storage…<br />
• Diagnostics… - Clicking this button downloads and<br />
displays a text file from the module. This file reports the<br />
performance of the module and connected devices. The<br />
measurement type will determine which information is<br />
included in this report. If you are not connected to the<br />
module, it displays the last available diagnostic report.<br />
• Hide in Measurements check box - when left unchecked,<br />
the Measurements tab displays this reading. If selected,<br />
the Measurements tab does not display this reading.<br />
If you have made any changes, click on the Apply button to<br />
update the settings in the module.<br />
See also:<br />
Level, Velocity, Flow Rate, Total Flow, Rainfall, Sampler, and<br />
Sonde tabs<br />
The Level tab of the Site window is used to view or change level<br />
measurement settings. Access this tab by clicking on the “Level”<br />
label on the Measurements tab.<br />
Modules able to read liquid levels can display this tab. From this<br />
tab you may:<br />
• View the Module and Measurement name.<br />
• View the current level measurement.<br />
56