<strong>Isco</strong> <strong>Flowlink</strong> <strong>5.1</strong> ® <strong>Software</strong><br />
Getting Started<br />
Connecting - 2100/<br />
4100/4200/6700<br />
Instruments<br />
Connecting - Pulse<br />
Doppler<br />
Instruments<br />
1. Click the Quick Connect button on the <strong>Flowlink</strong><br />
toolbar, or press F11 on the keyboard. <strong>Flowlink</strong> opens the<br />
Quick Connect window.<br />
2. Select the Type of connection- Direct, Modem, Wireless,<br />
or TCP.<br />
3. Check the Create new site check box and then click on<br />
the large button for the type of instrument that <strong>Flowlink</strong><br />
is to connect to.<br />
4. After <strong>Flowlink</strong> reads the instrument settings, it will<br />
pause and ask if you would like to clear the memory<br />
partitions (4100/4200/6700), or delete all of the data in<br />
the modules (2100). Select “No” to retain the data.<br />
<strong>Flowlink</strong> then completes the connection process.<br />
5. Once connected, <strong>Flowlink</strong> adds the instrument settings<br />
to its database and the site appears in the workspace.<br />
6. Click the Retrieve Data button. <strong>Flowlink</strong> downloads the<br />
stored readings and adds it to the database.<br />
NOTE: To retrieve only the data collected since the last<br />
interrogation, press Shift + F8 to open the Retrieve<br />
Options window.<br />
1. Click the Quick Connect button on the <strong>Flowlink</strong><br />
toolbar, or press F11 on the keyboard. <strong>Flowlink</strong> opens the<br />
Quick Connect window.<br />
2. Select the Type of connection- Direct, Modem, or<br />
Wireless.<br />
3. Check the Create new site check box and then click on<br />
the large button for pulse doppler instruments.<br />
The Pulse Doppler Connect window will appear.<br />
4. Enter a site name and a device name, and select the<br />
radio button for ADFM (Standard or HotTap), accQmin,<br />
or HADFM.<br />
Click OK, and <strong>Flowlink</strong> creates a site window.<br />
5. Click the Connect button, and <strong>Flowlink</strong> starts<br />
communication with the Pulse Doppler instrument.<br />
6. Once connected, <strong>Flowlink</strong> adds the instrument settings<br />
to its database and the site appears in the workspace.<br />
7. Click the Retrieve Data button. <strong>Flowlink</strong> downloads the<br />
stored readings and adds it to the database.<br />
8. NOTE: To retrieve only the data collected since the last<br />
interrogation, press Shift + F8 to open the Retrieve<br />
Options window.<br />
9