07.03.2014 Views

Flowlink 5.1 Software Instruction Manual - Isco

Flowlink 5.1 Software Instruction Manual - Isco

Flowlink 5.1 Software Instruction Manual - Isco

SHOW MORE
SHOW LESS

Create successful ePaper yourself

Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.

<strong>Isco</strong> <strong>Flowlink</strong> <strong>5.1</strong> ® <strong>Software</strong><br />

Editor Window<br />

251 Gap Record Used by DATAMAN, a <strong>Flowlink</strong> 2<br />

module, to indicate a gap in time<br />

between two merged data sets. This<br />

value appears only in readings<br />

imported from an edited <strong>Flowlink</strong> 2<br />

data set.<br />

250 Data Overflow The number of rainfall counts<br />

produced by the rain gauge<br />

exceeded the capacity of the<br />

memory location.<br />

248 Low Signal Strength Area velocity instruments use this<br />

No Data Code to indicate the flow<br />

stream does not contain enough<br />

reflective particles for the sensor to<br />

produce a velocity reading or that<br />

the sensor is covered with silt or<br />

sludge.<br />

247 Velocity Level Too<br />

Low<br />

Area velocity instruments use this<br />

No Data Code to indicate that the<br />

water level in the flow stream is<br />

below the minimum depth and the<br />

“set flow rate to zero if no velocity<br />

data” option is selected.<br />

141

Hooray! Your file is uploaded and ready to be published.

Saved successfully!

Ooh no, something went wrong!