Flowlink 5.1 Software Instruction Manual - Isco
Flowlink 5.1 Software Instruction Manual - Isco
Flowlink 5.1 Software Instruction Manual - Isco
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Isco</strong> <strong>Flowlink</strong> <strong>5.1</strong> ® <strong>Software</strong><br />
Site Window - Field Wizard<br />
Date/Time Box - displays the instrument’s and your computer’s<br />
time. These two times can be synchronized by pressing the<br />
“Synchronize Instrument’s Time To Computer’s” button. Use the<br />
Synchronize button with caution. Changing the time may affect<br />
the readings stored in the instrument’s memory partitions.<br />
Changing the time may cause readings to be overwritten, or<br />
“gaps” in readings may occur. It is recommended that you<br />
retrieve data from the instrument before synchronizing the time<br />
and date.<br />
If the Field Wizard is not in standalone mode, and is connected to<br />
<strong>Flowlink</strong> while part of a 2100 Series stack, the Instrument’s<br />
Time field will not be active.<br />
Connection<br />
Information Box<br />
Interrogated Sites -<br />
Field Wizard<br />
Field Wizard<br />
This box displays the settings used to communicate with the<br />
Field Wizard. When off-line you can modify:<br />
• Type - You should select the Direct option when<br />
connecting to the Field Wizard module.<br />
• COM port - Select the appropriate COM port setting<br />
from the pull-down menu. Refer to your computer’s<br />
documentation for more information on COM ports.<br />
• Baud Rate - Select the baud rate from the pull-down<br />
menu. Select 38400 when connecting to the Field Wizard.<br />
• Modem - (not used for the Field Wizard)<br />
• Phone number - (not used for the Field Wizard)<br />
To save changes to the communications settings, click Apply.<br />
The Interrogated Sites tab of the Field Wizard Site Window lists<br />
the sites that the Field Wizard interrogated and the modules<br />
that comprise each site.<br />
The top section of the display lists the interrogated site’s name,<br />
the date and time it was last interrogated, and whether the files<br />
are in a saved state.<br />
When you select a site name, the bottom section of the display<br />
will list the modules that exist at that site. When you select a<br />
module and click the Details button, it will display more details<br />
about that module.<br />
The 2101 Field Wizard is a portable data retrieval module<br />
designed to collect data from, or “interrogate” <strong>Isco</strong> 2100 Series<br />
flow modules. It then is able to transfer that data to <strong>Flowlink</strong>.<br />
The Field Wizard may also serve as a local display and keypad<br />
for a 2100 Series site.<br />
Use the Quick Connect window to begin communications<br />
between <strong>Flowlink</strong> and the Field Wizard.<br />
108