<strong>Isco</strong> <strong>Flowlink</strong> <strong>5.1</strong> ® <strong>Software</strong><br />

Site Window - Pulse Doppler<br />

• Baud Rate - (Direct) Select the baud rate of 38400 from<br />

the pull-down menu.<br />

• Modem - Select an installed modem from the pull-down<br />

list. <strong>Flowlink</strong> will only list those modems installed in<br />

Windows. <strong>Flowlink</strong> will use the selected Modem and its<br />

Windows settings (COM port, baud, etc.) to make the<br />

connection. If you need to change the modem settings,<br />

open the Modem Control Panel (Windows<br />

Start>Settings>Control Panel). A baud rate of 38400 is<br />

recommended for ADFM instruments.<br />

• Select Default from the list to use the modem<br />

specified in Utilities>Options for the appropriate<br />

device type. This automatically selects the modem<br />

used for a specific site, from any computer.<br />

• Phone number - (Modem) Enter the phone number<br />

assigned to the instrument. This number can include the<br />

following commands:<br />

T (tone dial)<br />

, (comma - pause)<br />

@ (wait for silence)<br />

W (wait for dial tone)<br />

• Pass-Through - (Interface Device or Service) From the<br />

pull down list, select the entity you are using to connect<br />

with the flow system.<br />

Battery Type and Projected Life - Select None if using an<br />

external power source, or Internal if using batteries.<br />

To save changes to the communications settings, click Apply.<br />

Measurements -<br />

Pulse Doppler<br />

When connected, <strong>Flowlink</strong> continually polls the site for “realtime”<br />

readings. This tab of the Site window displays the realtime<br />

readings, as well as test data, in the Output window.<br />

Program and Restart - writes your selected program settings<br />

to the instrument and starts the program.<br />

Erase Recorder - deletes all saved data from the internal<br />

recorder.<br />

Bench Test - The Bench Test function is for diagnostic purposes<br />

and is not required for routine operation. If for any reason the<br />

Bench Test is performed, regardless of the time interval between<br />

tests, first retrieve and save all recorded data before it is<br />

performed again. Prior to redeployment, select "Erase Recorder"<br />

and then "Restart."<br />

Field Test - This system performance evaluation is for use at the<br />

monitoring site with the sensor installed and in the stream.<br />

Quick Data - refreshes the Output window with the current<br />

data readings, and creates a graphical display to verify beam<br />

configuration, ensuring proper performance of the flow meter.<br />

Restart - restarts the instrument using the programmed<br />

parameters currently stored in its memory.<br />

103