Flowlink 5.1 Software Instruction Manual - Isco
Flowlink 5.1 Software Instruction Manual - Isco
Flowlink 5.1 Software Instruction Manual - Isco
You also want an ePaper? Increase the reach of your titles
YUMPU automatically turns print PDFs into web optimized ePapers that Google loves.
<strong>Isco</strong> <strong>Flowlink</strong> <strong>5.1</strong> ® <strong>Software</strong><br />
Site Window - 4100/4200/6700<br />
Time Table<br />
Conditions<br />
Parentheses and<br />
Order of Execution<br />
Time table conditions are satisfied when the instrument’s<br />
internal clock is within the defined duration of time. An example<br />
of a Daily Time table condition would read “starting Daily at<br />
08:00:00, stopping Daily at 17:00:00.”<br />
As many as four times can be entered, using any combination of<br />
time table types. For example, to enable a function every Monday,<br />
Wednesday, and Friday, three weekly start/stop times would be<br />
defined in the time table. The three lines would possibly read<br />
“Starting Monday at 08:00:00, Stopping at Monday 17:00:00;<br />
Starting Wednesday at 08:00:00, Stopping Wednesday at<br />
17:00:00; and Starting Friday at 08:00:00, Stopping Friday at<br />
17:00:00.”<br />
To define the time table<br />
1. Select the type of time table. The options are:<br />
• Specific – Start/stop dates and times for one-time<br />
use.<br />
• Weekly – Start/stop days of week and times,<br />
recurring weekly.<br />
• Daily – Start/stop times, recurring daily.<br />
• Mon-Fri – Start/stop times, recurring each day of the<br />
workweek.<br />
2. Complete the Start and Stop fields. The resulting<br />
duration will be displayed in the text box below the entry<br />
fields.<br />
3. To define additional start/stop times, click Add and<br />
repeat the steps above.<br />
Note: Start/stop times can be deleted by highlighting the<br />
line in the text box and clicking Remove.<br />
4. Click Accept. The time table is defined and ready for use<br />
in an equation.<br />
<strong>Flowlink</strong> provides parentheses so that you can group conditions<br />
or even combine groups with logical operators. Parentheses also<br />
let you control the order in which the flow meter evaluates the<br />
conditions in an equation. The instrument will use an algebraic<br />
order of execution when evaluating a trigger equation. The order<br />
of execution is the order in which the instrument tests conditions<br />
and can be summarized as: “from left to right, and from inside,<br />
out.” “From left to right” means that the flow meter tests<br />
conditions in the same order that you read them in an equation:<br />
from left to right.<br />
When evaluating an equation containing two conditions, the<br />
instrument first checks the condition on the left, then the<br />
condition on the right, and finally evaluates the conditions with<br />
regard to the logical operators used to combine them. If the<br />
equation is true, the instrument triggers the designated action.<br />
Equations using three or more conditions are also evaluated from<br />
left to right. Consider the two equations that follow. Both<br />
97