<strong>Isco</strong> <strong>Flowlink</strong> <strong>5.1</strong> ® <strong>Software</strong><br />

Site Window - 4100/4200/6700<br />

The On trigger setting requires a trigger to be set up.<br />

Pressing the Change Enable button opens the Equation<br />

Builder to set up the trigger conditions. Once this has<br />

been entered, the trigger is displayed in the “Enable<br />

sampler when:” box.<br />

Latch sampler enable - If this box is unchecked, the inhibit<br />

line will continue to toggle between enabled and disabled as the<br />

conditions change. When this box is checked, the inhibit line will<br />

keep the sampler enabled once triggered, even if the condition no<br />

longer exists. This state will not change until the Reset Latch<br />

button is pressed.<br />

Disable plotter when not sampling - When this box is<br />

checked, the instrument’s plotter will only run when the sampler<br />

is enabled.<br />

Block access to front panel - When this box is checked, front<br />

panel sampler enable changes are not possible and the flow<br />

meter displays a message stating that the access has been<br />

blocked.<br />

Sampler Pacing<br />

Flow Pulse<br />

Control the sampler pacing by choosing one of the pacing options.<br />

• No sampler pacing - The flow logger or flow meter will<br />

not send flow pulses.<br />

• Flow paced - paces the sampler at a rate proportional to<br />

the flow rate by using volume intervals. Use the Change<br />

Pacing button to set the interval.<br />

• Trigger paced - paces the sampler at a time interval that<br />

can change according to a triggered condition.<br />

To set up trigger paced sampling:<br />

1. Select the Trigger paced radio button. The Time paced<br />

interval box becomes active.<br />

2. Enter the When trigger is true interval in minutes. This<br />

is the pacing interval while the trigger condition is met.<br />

3. Enter the When trigger is false interval in minutes. This<br />

is the pacing interval while the trigger condition has not<br />

been met.<br />

4. Set the trigger by pressing the Change Pacing button.<br />

Build the trigger in the Equation Builder.<br />

5. Press the Apply button.<br />

Use this window to set the flow pulse interval. A flow pulse is a<br />

50 millisecond pulse used to pace a sampler. Enter a value within<br />

the minimum and maximum range listed.<br />

It is important that this setting and the sampler’s pulses<br />

between sampling events setting work together to produce the<br />

desired volume interval. For example, if the flow meter sends a<br />

flow pulse every 100 gallons, and the sampler takes a sample<br />

every 10 flow pulses, the actual volume between samples is 1000<br />

gallons.<br />

91