RediSep Alumina Neutral 80g Column Care Guide - Isco
RediSep Alumina Neutral 80g Column Care Guide - Isco
RediSep Alumina Neutral 80g Column Care Guide - Isco
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
<strong>RediSep</strong> ® <strong>Alumina</strong> <strong>Neutral</strong><br />
Functionalized <strong>Column</strong>, <strong>80g</strong><br />
<strong>Column</strong> <strong>Care</strong> Instructions<br />
Instruction Sheet 60-2203-237<br />
<strong>Column</strong> Information and Sample Run<br />
Proper conditioning and use of Teledyne <strong>Isco</strong>’s<br />
<strong>RediSep</strong> ® <strong>Alumina</strong> <strong>Neutral</strong> functionalized columns will<br />
help to improve column consistency and reliability. This<br />
instruction sheet provides a sample run that may be<br />
used as a column performance benchmark.<br />
<strong>Column</strong> <strong>Care</strong><br />
Initial <strong>Column</strong> Conditioning<br />
The column should be equilibrated with a minimum<br />
of three column volumes. Normally columns are equilibrated<br />
with the initial conditions of the gradient. The<br />
equilibration can be written into the method if desired.<br />
Equilibration Between Runs<br />
<strong>RediSep</strong> ® <strong>Alumina</strong> <strong>Neutral</strong> columns are intended for<br />
single-use only. However, repeat use of the column may<br />
be possible assuming that all sample was eluted off the<br />
column during the previous run. Three to five column<br />
volumes are a sufficient amount of equilibration<br />
between runs. If baseline drift is noticeable early in the<br />
run then more equilibration may be needed between<br />
runs.<br />
Sample Run<br />
The sample run purifies a 1:1 mixture of 6-Chlorochromone<br />
and Propyl 4-hydroxybenzoate (Figure 1)<br />
using the parameters in Table 1. The chromatogram is<br />
illustrated in Figure 2.<br />
Table 1: General Method Parameters<br />
<strong>Column</strong>:<br />
Wavelength:<br />
Sample loading method:<br />
Mobile phase:<br />
Flow Rate:<br />
Equilibration Volume:<br />
Gradient:<br />
Cl<br />
Figure 1: Compound mixture<br />
<strong>RediSep</strong> <strong>80g</strong> <strong>Alumina</strong> <strong>Neutral</strong> Functionalized<br />
<strong>Column</strong> (68-2203-087)<br />
254 nm<br />
624 mg pre-loaded on silica gel<br />
Solvent A: Hexane<br />
Solvent B: Ethyl Acetate<br />
40 mL/minute<br />
4 <strong>Column</strong> Volumes<br />
% Solvent B<br />
0<br />
0<br />
100<br />
100<br />
0<br />
0<br />
O<br />
6-Chlorochromone<br />
O<br />
OH<br />
Minutes<br />
Initial<br />
2.9<br />
41.3<br />
8.8<br />
0.0<br />
2.9<br />
O<br />
Propyl 4-hydroxybenzoate<br />
Assistance Available<br />
Should you need assistance with these instructions,<br />
please contact Teledyne <strong>Isco</strong>.
Instruction Sheet 60-2203-237<br />
Propyl 4-hydroxybenzoate<br />
UV Absorbance<br />
6-Chlorochromone<br />
%B Mobile Phase<br />
Figure 2: Sample Run Chromatogram<br />
Minutes<br />
Last modified October 19, 2004<br />
Teledyne <strong>Isco</strong>, Inc.<br />
P.O. Box 82531, Lincoln, Nebraska, 68501 USA<br />
Toll-free: (800) 228-4373 • Phone: (402) 464-0231 • Fax: (402) 465-3091<br />
E-mail: info@isco.com