<strong>RediSep</strong> ® <strong>Alumina</strong> <strong>Neutral</strong><br />

Functionalized <strong>Column</strong>, <strong>80g</strong><br />

<strong>Column</strong> <strong>Care</strong> Instructions<br />

Instruction Sheet 60-2203-237<br />

<strong>Column</strong> Information and Sample Run<br />

Proper conditioning and use of Teledyne <strong>Isco</strong>’s<br />

<strong>RediSep</strong> ® <strong>Alumina</strong> <strong>Neutral</strong> functionalized columns will<br />

help to improve column consistency and reliability. This<br />

instruction sheet provides a sample run that may be<br />

used as a column performance benchmark.<br />

<strong>Column</strong> <strong>Care</strong><br />

Initial <strong>Column</strong> Conditioning<br />

The column should be equilibrated with a minimum<br />

of three column volumes. Normally columns are equilibrated<br />

with the initial conditions of the gradient. The<br />

equilibration can be written into the method if desired.<br />

Equilibration Between Runs<br />

<strong>RediSep</strong> ® <strong>Alumina</strong> <strong>Neutral</strong> columns are intended for<br />

single-use only. However, repeat use of the column may<br />

be possible assuming that all sample was eluted off the<br />

column during the previous run. Three to five column<br />

volumes are a sufficient amount of equilibration<br />

between runs. If baseline drift is noticeable early in the<br />

run then more equilibration may be needed between<br />

runs.<br />

Sample Run<br />

The sample run purifies a 1:1 mixture of 6-Chlorochromone<br />

and Propyl 4-hydroxybenzoate (Figure 1)<br />

using the parameters in Table 1. The chromatogram is<br />

illustrated in Figure 2.<br />

Table 1: General Method Parameters<br />

<strong>Column</strong>:<br />

Wavelength:<br />

Sample loading method:<br />

Mobile phase:<br />

Flow Rate:<br />

Equilibration Volume:<br />

Gradient:<br />

Cl<br />

Figure 1: Compound mixture<br />

<strong>RediSep</strong> <strong>80g</strong> <strong>Alumina</strong> <strong>Neutral</strong> Functionalized<br />

<strong>Column</strong> (68-2203-087)<br />

254 nm<br />

624 mg pre-loaded on silica gel<br />

Solvent A: Hexane<br />

Solvent B: Ethyl Acetate<br />

40 mL/minute<br />

4 <strong>Column</strong> Volumes<br />

% Solvent B<br />

0<br />

0<br />

100<br />

100<br />

0<br />

0<br />

O<br />

6-Chlorochromone<br />

O<br />

OH<br />

Minutes<br />

Initial<br />

2.9<br />

41.3<br />

8.8<br />

0.0<br />

2.9<br />

O<br />

Propyl 4-hydroxybenzoate<br />

Assistance Available<br />

Should you need assistance with these instructions,<br />

please contact Teledyne <strong>Isco</strong>.


Instruction Sheet 60-2203-237<br />

Propyl 4-hydroxybenzoate<br />

UV Absorbance<br />

6-Chlorochromone<br />

%B Mobile Phase<br />

Figure 2: Sample Run Chromatogram<br />

Minutes<br />

Last modified October 19, 2004<br />

Teledyne <strong>Isco</strong>, Inc.<br />

P.O. Box 82531, Lincoln, Nebraska, 68501 USA<br />

Toll-free: (800) 228-4373 • Phone: (402) 464-0231 • Fax: (402) 465-3091<br />

E-mail: info@isco.com