chamber and the freezer, only one working at a t<strong>im</strong>e. An independent temperature<br />

sensor connected to the main security switch was applied to disconnect all power<br />

devices in the system in case of reaching upper l<strong>im</strong>it temperature, Tz = 65 ºC,<br />

inside the chamber.<br />

Fig. 19.<br />

Block diagram of the laboratory setup for the determination of temperature effect on soil<br />

dielectric permittivity<br />

Each TDR measurement started when the air temperature inside the chamber,<br />

Tp, the temperature of the measured soil sample, T0, T1, .. or T7, and the<br />

regulation temperature reached the same value with ±0.4 ºC tolerance. The<br />

decrease of the temperature tolerance was possible but it would cost the<br />

significant increase of t<strong>im</strong>e to stabilize the temperature in the soil sample and the<br />

chamber. The complete measurement cycle at six temperature values, from 5 ºC<br />

to 55 ºC in 10 ºC increments, for the set of eight soil samples took about 12 hours<br />

and starting from the 4-th cycle the measurable decrease of TDR water content<br />

45