Measurement Functions<br />
Zoom (Y-Axis): The scaling of the spectrum’s Y-axis (sensitivity) is<br />
fixed and may be adjusted manually.<br />
• Select the zoom (Y-axis) field and press the enter key.<br />
• Use the up/down keys to scroll the displayed level along the Yaxis<br />
and the left/right keys zoom out/in the Y-axis, e.g. alter the<br />
resolution of the division.<br />
Time Constant: The time constant corresponds to the integration time<br />
of the 1/3 rd octave recording. It may be set to five value: 0.2 / 0.5 /<br />
1.0 / 2.0 / 5.0 seconds. A lower time constant results in a quicker,<br />
more ‘nervous' display, whilst a higher time constant ‘averages’<br />
more samples, thus providing a more stable spectrum.<br />
Bargraph: Below the spectrum, the fullband input RMS level (20 Hz<br />
- 20 kHz) is indicated by a bargraph without auto ranging. Select the<br />
different indication ranges by moving the cursor to the left or right<br />
arrow below RNGE and press enter. The available ranges are:<br />
• -120 to -20 dBu, residual noise of <strong>ML1</strong> < -120dBu (