AxSOS Locking Plate System - Stryker
AxSOS Locking Plate System - Stryker
AxSOS Locking Plate System - Stryker
Create successful ePaper yourself
Turn your PDF publications into a flip-book with our unique Google optimized e-Paper software.
Operative Technique<br />
Percutaneous Screw Insertion<br />
A small stab incision can then be made<br />
through the slot to locate the hole<br />
selected for screw placement. The Shaft<br />
Hole Locator can then be rotated out<br />
of the way or removed.<br />
The Standard Percutaneous Drill<br />
Sleeve (REF. 702710) or the Neutral<br />
Percutaneous Drill Sleeve<br />
(REF 702958) in conjunction with<br />
the Drill Sleeve Handle (REF 702822)<br />
can be used to assist with drilling for<br />
Standard Screws. Use a 3.2mm<br />
drill bit (REF 700357).<br />
With the aid of the Soft Tissue<br />
Spreader (REF 702918), the skin can<br />
be opened to form a small window<br />
(Fig. 19–20) through which either<br />
a Standard Screw or <strong>Locking</strong> Screw<br />
(provided a <strong>Locking</strong> Insert is present)<br />
can be placed. For <strong>Locking</strong> Screw<br />
insertion, use the threaded Drill<br />
Sleeve (REF 702708) together with<br />
the 4.3mm drill bit (REF 702743)<br />
to drill the core hole.<br />
Fig. 19<br />
Fig. 20<br />
15