110.0<br />

CLIO<br />

180.0<br />

dBSPL<br />

Deg<br />

100.0 108.0<br />

90.0 36.0<br />

80.0 -36.0<br />

70.0 -108.0<br />

60.0 -180.0<br />

20 100 1k Hz<br />

10k 20k<br />

Figure 10.27<br />

To terminate this difficult paragraph we will summarise what we did with some<br />

comments. Measuring acoustic phase response is often far from a "press a button and<br />

get it" procedure. We went trough several phase plots, all looking different but, this is<br />

an important point, all correct. It is common to identify the processed phase as the true<br />

one only because it looks better. It is important to stress that the true phase is that of<br />

Figures 10.18 and 10.19. CLIO, intended as a computer based instrument, can, as we<br />

will see later, easily calculate the summed response of woofer and tweeter after their<br />

are taken separately but with the same microphone position. What we did with our<br />

complicated procedure was to obtain a response as if the microphone would have been<br />

exactly in the acoustic centre of the driver. The most obvious application is to furnish<br />

data to cross-over CAD.<br />

Chapter 10 - MLS 99