Figure 10.27<br />
To terminate this difficult paragraph we will summarise what we did with some<br />
comments. Measuring acoustic phase response is often far from a "press a button and<br />
get it" procedure. We went trough several phase plots, all looking different but, this is<br />
an important point, all correct. It is common to identify the processed phase as the true<br />
one only because it looks better. It is important to stress that the true phase is that of<br />
Figures 10.18 and 10.19. CLIO, intended as a computer based instrument, can, as we<br />
will see later, easily calculate the summed response of woofer and tweeter after their<br />
are taken separately but with the same microphone position. What we did with our<br />
complicated procedure was to obtain a response as if the microphone would have been<br />
exactly in the acoustic centre of the driver. The most obvious application is to furnish<br />
data to cross-over CAD.<br />
Chapter 10 - MLS 99