6.3.5 CALIBRATION<br />

File > Calibration<br />

This option will perform a calibration of your CLIO hardware. Please refer to paragraph<br />

3.7 and follow the procedure described.<br />

In order to determine, at any given time, if it is necessary to calibrate CLIO do the<br />

following:<br />

- Let the system warm up<br />

- Proceed to perform the verification described in 3.7.1<br />

- Consequently decide whether or not to calibrate<br />

The result of the measurement may vary in some way from the time we calibrated<br />

because of many small changes in measurement conditions, including changes in the<br />

atmospheric conditions, the season, and the mains voltage.<br />

Note: the CLIO hardware is highly precise and stable and, under normal<br />

operating conditions, does not require frequent calibrations.<br />

Always perform a calibration if:<br />

- You reinstalled CLIO in a different computer<br />

- You installed a software upgrade<br />

6.3.6 SETUP<br />

File > Setup<br />

This option gives you access to the setup dialog (Fig. 6.7), where you can define the<br />

following:<br />

- Screen Colors<br />

- Print (and graphics export) colors<br />

- Screen line width<br />

- Print (and graphics export) line width<br />

- The Company Name which will appear in all printouts.<br />

Fig.6.7 Setup dialog<br />

62 Chapter 6 - System Operations and Settings